Physiological effects of different interfaces during noninvasive ventilation for acute respiratory failure.

Fraticelli Amanda Tarabini, Lellouche François, L'her Erwan, Taillé Solenne, Mancebo Jordi, Brochard Laurent

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To test the short-term physiologic effects (indexes of respiratory effort, ventilation, and gas exchange), leaks, patient-ventilator asynchrony, and comfort of four noninvasive ventilation (NIV) facial, oronasal, or oral interfaces with major differences in internal volume.

DESIGN

Prospective, short-term, crossover randomized physiologic study.

SETTING

Medical intensive care unit in a university hospital.

PATIENTS AND

PARTICIPANTS

Fourteen consecutive patients receiving NIV for either hypoxemic (n = 7) or hypercapnic (n = 7) acute respiratory failure.

INTERVENTIONS: Four interfaces, tested randomly over consecutive sequences, had very high (977 mL), high (163 mL), moderate (84 mL), or virtually no internal volume (mouthpiece). The pressure level was increased in two patients with the larger mask, and was decreased in all patients when using the mouthpiece.

MEASUREMENTS AND MAIN

RESULTS

Despite differences in internal volume, no apparent dead space effect was observed on minute ventilation, work of breathing, or arterial CO2 levels. Compared with baseline, NIV was uniformly successful in reducing indexes of respiratory effort: the pressure-time product of the respiratory muscles decreased from a median (25th-75th interquartile range) of 179 (158-285) cm H2O.sec.min to values between 91 and 111 during NIV, with no differences between masks (p = 0.84). Few differences were observed in terms of arterial blood gases and breathing pattern. Leaks and patient-ventilator asynchronies were greater with the mouthpiece, and comfort with this interface was deemed poor for most patients.

CONCLUSION

The internal volume of the masks had no apparent short-term dead space effect on gas exchange, minute ventilation, or patient's effort, suggesting that, with the exception of mouthpiece, interfaces may be interchangeable in clinical practice provided adjustment of the ventilatory device parameters are performed.

Для цитирования: Fraticelli Amanda Tarabini, Lellouche François, L'her Erwan, Taillé Solenne, Mancebo Jordi, Brochard Laurent. Physiological effects of different interfaces during noninvasive ventilation for acute respiratory failure.. Critical care medicine. 2009. DOI: 10.1097/ccm.0b013e31819b575f.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.