DOI 10.1016/j.ajem.2007.04.022
An observational study of noninvasive positive pressure ventilation in an out-of-hospital setting.
Аннотация
OBJECTIVE
Out-of-hospital clinical experience with noninvasive bilevel positive airway pressure (BiPAP) ventilation is extremely limited compared to inhospital management. The aims of this study were to assess the feasibility of out-of-hospital BiPAP ventilation in patients with acute respiratory distress of various origins, and to look for specific factors associated with failure of this respiratory support.
METHODS
This 2-year prospective observational study assessed the failure rate of out-of-hospital BiPAP ventilation, the difficulties encountered, and factors predictive of failure by multivariate analysis.
RESULTS
Overall, 138 patients were treated by out-of-hospital BiPAP for congestive heart failure (56%), chronic obstructive pulmonary disease exacerbation (28%), and acute respiratory failure (16%). Failure rate was 26% (35/138; 95% confidence interval, 18%-33%) (11 before and 24 after reaching hospital). Independent risk factors were the cause of respiratory distress (chronic obstructive pulmonary disease exacerbation; acute respiratory failure) and an audible air leakage.
CONCLUSIONS
The failure rate of BiPAP initiated out-of-hospital was no different from previous reports for inhospital failure rates. Failure was attributable to similar causes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.