DOI 10.1111/vsu.14328

Медицинская библиотека · Veterinary surgery : VS · 2025

Biomechanical testing of three constructs for prosthetic laryngoplasty in horses demonstrates advantages of differing metallic implants in the arytenoid cartilage.

Ysebaert Machiel P, Johnson James, Marie Ulrika, Campos Arian, Verchrerin Anouk, Ducharme Norm G, Rossignol Fabrice, Luedke Lauren K

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the biomechanical stability of a novel arytenoid implant (tie-bolt, TB) with a suture anchor (SA) and standard laryngoplasty using suture alone (control) in equine larynges.

STUDY

DESIGN

Randomized ex vivo controlled experimental study.

SAMPLE POPULATION: Thirty equine cadaver larynges.

METHODS

Larynges were randomly assigned to TB, SA, or control groups (n = 10 each). The force of abduction was measured for each construct, then constructs underwent cyclic loading (5-50 N for 3000 cycles) to assess elongation, followed by load-to-failure testing to assess ultimate load, displacement, stiffness, and mode of failure.

RESULTS

The SA required less force to achieve 88% abduction (8.31 ± 2.46 N) than the TB (11.48 ± 3.6 N, p = .07) and controls (11.86 ± 2.99 N, p = .04). The TB had less construct elongation (2.62 ± 0.61 mm) than SA constructs (3.87 ± 1.25 mm, p = .01) and was stiffer (31.05 ± 3.38 N/mm) than both control (23.48 ± 2.06 N/mm, p < .0001) and SA constructs (18.53 ± 4.15 N/mm, p < .0001). The TB construct failed at higher ultimate load and displacement (367.25 ± 66.98 N and 19.11 ± 2.51 mm, respectively) than the control (238.14 ± 19.46 N and 14.23 ± 1.6 mm; p < .0001 for both) and SA constructs (196.55 ± 47.17 N and 16.23 ± 2.02 mm; p < .0001 and p = .01, respectively). The primary modes of failure were suture cut-through for control, implant pullout for SA, and arytenoid fracture for TB.

CONCLUSION

The SA required the least force for abduction. The TB was a stiffer construct that resulted in higher failure loads.

CLINICAL SIGNIFICANCE: The SA may reduce construct loading during abduction. The TB offers superior mechanical stability, potentially reducing long-term degradation from cyclic loading.

Для цитирования: Ysebaert Machiel P, Johnson James, Marie Ulrika, Campos Arian, Verchrerin Anouk, Ducharme Norm G, Rossignol Fabrice, Luedke Lauren K. Biomechanical testing of three constructs for prosthetic laryngoplasty in horses demonstrates advantages of differing metallic implants in the arytenoid cartilage.. Veterinary surgery : VS. 2025. DOI: 10.1111/vsu.14328.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.