Early rupture of a tetrafluoroethylene loop due to fixation with polypropylene resulting in recurrent mitral valve regurgitation; a word of caution.

Mitsube Keijiro, Wakabayashi Naohiro, Tsutsui Masahiro, Ushioda Ryohei, Kamiya Hiroyuki, Fujita Tsutomu

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

This case report describes early rupture of an expanded polytetrafluoroethylene (ePTFE) loop used in mitral valve repair, likely due to leaflet fixation with 5/0 polypropylene suture. A 52-year-old man developed recurrent mitral valve regurgitation 3 weeks postoperatively. Reoperation revealed rupture at the fixation site. This case suggests that ePTFE may be a more suitable material than polypropylene for leaflet fixation to prevent early loop failure.

Для цитирования: Mitsube Keijiro, Wakabayashi Naohiro, Tsutsui Masahiro, Ushioda Ryohei, Kamiya Hiroyuki, Fujita Tsutomu. Early rupture of a tetrafluoroethylene loop due to fixation with polypropylene resulting in recurrent mitral valve regurgitation; a word of caution.. Journal of surgical case reports. 2025. DOI: 10.1093/jscr/rjaf393.