Медицинская библиотека · RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 1996

[Temporary vena cave filter: critical comments].

Textor H J, Strunk H, Schild H H

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Retrospective evaluation of the reasons for complications caused by the placement of temporary vena cava filters.

METHOD: Retrospective analysis of 9 complications in 6 implanted temporary (2 anthéor TC-, 4 Filcard RF 02-Filter) and 6 retrievable vena cava filters (1 Dil-, 5 FCP 2002-Filter).

RESULTS

We observed 7 filter thromboses and two severe infections. 5 filter thromboses occurred in temporary filters, two filter thromboses and two infections occurred in facultative temporary filters.

CONCLUSION

The main complications of temporary vena cava filters, like filter thrombosis and infection, were associated with the fixation mechanism of the filter. Our results indicate that facultative temporary vena cava filters without a permanent percutaneous retrieval system seem to be the safer system.

Для цитирования: Textor H J, Strunk H, Schild H H. [Temporary vena cave filter: critical comments].. RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 1996. DOI: 10.1055/s-2007-1015771.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.