A correction to the bandwidth difference method for assessment of magnetic resonance static field homogeneity.

Wishart Joseph, Brunnquell Christina L, Vergara Daniel, Ostenson Jason

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Published guidance offers four methods of static field inhomogeneity ( Δ B 0 ) measurement, including a bandwidth difference method. However, an error exists in the implementation of this method, which we demonstrate analytically and experimentally.

PURPOSE: To propose and validate a correction to the bandwidth difference method for quantifying Δ B 0 .

METHODS

We outline a correction in the method for quantifying local Δ B 0 by measuring spatial translations rather than diameter measurements between MRI images with different receive bandwidths. To validate the approach, we acquired images of a cylindrical phantom containing a fixed array of fluid-filled pins. We adjusted static field shims to create approximately linear and quadratic field perturbations and acquired high- and low-bandwidth spoiled gradient echo sequences with anterior-posterior (AP) and left-right (LR) frequency-encoding. We then measured translations of centroid pin locations between low- and high-bandwidth images. Finally, we propose and validate an implementation strategy using a spherical phantom. In both cases, a vendor-provided field mapping sequence provided reference homogeneity measurements.

RESULTS

Under manually adjusted shim settings, root-mean-square (RMS) and peak-to-peak (PP) Δ B 0 were 2.46 and 6.29 parts per million (PPM) respectively, measured with the field mapping technique. Results of the proposed bandwidth difference method showed strong agreement; RMS and PP Δ B 0 were 2.38 and 6.22 PPM respectively for AP frequency-encoding and 2.38 and 6.08 PPM respectively for LR frequency-encoding. In a spherical phantom, the proposed method agreed with reference measurements to within 0.25 PPM (< 7%). However, PP Δ B 0 was underestimated using the diameter-based method-in a spherical phantom, measured Δ B 0 was 0.74 PPM compared to 3.52 and 3.76 PPM using the reference and revised methods, respectively.

CONCLUSIONS

This work experimentally validates a revised bandwidth difference approach for Δ B 0 quantification, showing that it remains sensitive under spatially nonlinear Δ B 0 ; the diameter-based method is shown to be inaccurate for such cases.

Для цитирования: Wishart Joseph, Brunnquell Christina L, Vergara Daniel, Ostenson Jason. A correction to the bandwidth difference method for assessment of magnetic resonance static field homogeneity.. Medical physics. 2026. DOI: 10.1002/mp.70673.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.