Lot variation and inter-device differences contribute to poor analytical performance of the DCA Vantage™ HbA1c POCT instrument in a true clinical setting.
Аннотация
OBJECTIVES
The glycated haemoglobin fraction A1c (HbA1c) is widely used in the management of diabetes mellitus, and the Siemens DCA Vantage™ point-of-care testing (POCT) instrument offers rapid HbA1c results even far from a clinical laboratory. However, the analytical performance has been questioned, and not much is known about effects of changing reagent lot, instrument and operator. We therefore compared the analytical performance of the DCA Vantage™ with established routine methods (Tosoh G8/G11 ion exchange HPLC) in a true clinical setting at two Danish hospitals.
METHODS
We extracted all routine clinical HbA1c results incidentally drawn from the same patient within 48 h (n=960 pairs) and evaluated the effect of reagent lot, operator and instrument. We also performed a prospective method comparison in our diabetes out-patient clinic (n=97).
RESULTS
The critical difference (CD) between two POCT results varied between 5.14 and 6.61 mmol/mol (0.47-0.55%), and the analytical imprecision of the DCA Vantage™ (CVA) was >3%. Significant effect of reagent lot and inter-instrument differences were found, whereas no effect of operator was seen.
CONCLUSIONS
The DCA Vantage™ HbA1c analysis does not fulfil the prevailing analytical performance specifications, but rigorous validation of new reagent lots and continuous recalibration of instruments may potentially improve the precision substantially. Our findings, therefore, clearly emphasise the necessity of a close collaboration between clinicians and laboratory professionals in the POCT field. Finally, POCT HbA1c results should always be interpreted together with other measures of glycaemic control to avoid inappropriate change of patient treatments due to measurement uncertainty.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.