Ineffective cardiac resynchronization pacing is associated with poor outcomes in a nationwide cohort analysis.

Robbins-Juarez Shelief Y, Mittal Suneet, Plummer Chris, Koehler Jodi L, Stadler Robert W, Ghosh Subham, Klepfer Ruth N, Piccini Jonathan P

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Delivery of cardiac resynchronization therapy (CRT) requires left ventricular myocardial capture to achieve clinical benefits.

OBJECTIVE

We sought to determine whether ineffective pacing affects survival.

METHODS

Ineffective ventricular pacing (VP) was defined as the difference between the percentage of delivered CRT (%VP) and the percentage of effective CRT in CRT devices. Using the Optum de-identified electronic health record data set and Medtronic CareLink data warehouse, we identified patients implanted with applicable devices with at least 30 days of follow-up. Kaplan-Meier and Cox proportional hazards models assessed the impact of %VP and % ineffective VP on survival.

RESULTS

Among 7987 patients with 2.1 ± 1.0 years of follow-up, increasing ineffective VP was associated with decreasing survival: the highest observed survival was in the quartile with <0.08% ineffective VP and the lowest survival was in the quartile with >1.47% ineffective VP (85.1% vs 75.7% at 3 years; P < .001). As expected, patients with more than the median %VP of 97.7% had better survival than did patients with <97.7% VP (84.2% vs 77.8%; P < .001). However, patients who had >97.7% VP but >2% ineffective VP had similar survival to patients with <97.7% VP but ≤2% ineffective VP (81.6% vs 79.4%; P = .54). A multivariable Cox proportional hazards model demonstrated that <97.7% VP (adjusted hazard ratio 1.29; 95% confidence interval 1.14-1.46; P < .001) and >2% ineffective VP (hazard ratio 1.35; 95% confidence interval 1.18-1.54; P < .001) were both significantly associated with decreased survival.

CONCLUSION

Ineffective VP is associated with decreased survival. In addition to maximizing the percentage of delivered CRT pacing, every effort should be made to minimize ineffective VP.

Для цитирования: Robbins-Juarez Shelief Y, Mittal Suneet, Plummer Chris, Koehler Jodi L, Stadler Robert W, Ghosh Subham, Klepfer Ruth N, Piccini Jonathan P. Ineffective cardiac resynchronization pacing is associated with poor outcomes in a nationwide cohort analysis.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.06.064.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.