Continuous Glucose Monitoring System After Coronary Artery Bypass Graft Surgery: A Feasibility Study.
Аннотация
BACKGROUND
Blood glucose level variability has been associated with increased risk of complication in the postoperative setting of cardiovascular surgery. Although interesting for optimization of blood glucose management in this context, continuous blood glucose (CBG) devices can have a limited reliability in this context, in particular because of the use of paracetamol. The aim of this study was to evaluate the reliability of Dexcom G6®, a recently developed continuous glucose monitoring device.
METHODS
We performed a prospective, observational, non-randomized, single-centre study comparing Dexcom G6® CBG level monitoring with the standard methods routinely used in this context. The standard blood glucose values were paired to the time corresponding values measured with Dexcom G6®. Agreement between the two methods and potential correlation in case of paracetamol use were calculated.
RESULTS
From May 2020 to August 2021, 36 out of 206 patients operated for isolated coronary artery bypass grafting were enrolled; 673 paired blood glucose level were analyzed. Global agreement (ρc) was 0.85 (95% C.I.: 0.84-0.86), intensive care unit agreement was 0.78 (95%C.I.: 0.74-0.82) and ward agreement was 0.91 (95%C.I.: 0.89-0.93). In the diabetic population, it was 0.87 (95%C.I.: 0.85-0.90). When paracetamol was used, the difference was 0.02 mmol/l (95%C.I.: 0.29-0.33).
CONCLUSIONS
Dexcom G6® provides good blood glucose level accuracy in the postoperative context of cardiac surgery compared to the standard methods of measurements. The results are particularly reliable in the ward where the need for repeated capillary glucose measurements implies patient discomfort and time-consuming manipulations for the nursing staff.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.