DOI 10.25270/jic/24.00011

Медицинская библиотека · The Journal of invasive cardiology · 2024

Optimal oversizing in transcatheter aortic valve replacement with the self-expanding Evolut valve system.

Mas-Peiro Silvia, Alperi Alberto, Avvedimento Marisa, Regueiro Ander, Avanzas Pablo, Angellotti Domenico, Esposito Giovanni, Vidal-Cales Pablo, Mohammadi Siamak, Farjat-Pasos Julio, Moris Cesar, Rodés-Cabau Josep

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Valve oversizing has been associated with reduced paravalvular leaks (PVL) and valve migration risk. However, no optimal cut-off oversizing value has been defined for the Evolut system (Medtronic). The aim of this study was to assess the relationship between the degree of oversizing and moderate-to-severe PVL and determine the optimal oversizing cut-off value.

METHODS

We conducted a multicenter study that included 740 consecutive patients with multidetector computed tomography (CT) data. Valve size was selected according to manufacturer recommendations, with oversizing ranging from 10% to 30%. The primary endpoint was moderate-to-severe PVL.

RESULTS

The median age was 84 years (79-87 years), with 58.4% women, and a median EuroSCORE II of 4.1% (2.4-7.3%). Moderate-to-severe PVL was observed in 7.0% of the patients. An inverse relationship was found between oversizing and both PVL (11.3%, 8.6%, 5.4%, and 2.7% for quartiles Q1 to Q4; P = .007) and the need for post-dilation (P = .016). The multivariable analysis showed an association between oversizing and PVL (OR: 0.915 for each 1%-increase, P = .002). The optimal oversizing cut-off value to predict PVL was 20%, and PVL was significantly higher in patients with oversizing less than 20% (10.5% vs.4.2%, P less than .001). There were no differences in major clinical events according to the degree of oversizing, and a higher oversizing did not translate into an increased risk of permanent pacemaker (18.4% vs18.3%, P = .976).

CONCLUSIONS

In TAVR with the Evolut valve, a higher oversizing was associated with lower rates of moderate-to-severe PVL and a lower need for post-dilation, with no negative impact on procedural and early clinical outcomes. A 20% oversizing threshold could be suggested to reduce PVLs. Further prospective studies are warranted to validate optimal oversizing for this valve system.

Для цитирования: Mas-Peiro Silvia, Alperi Alberto, Avvedimento Marisa, Regueiro Ander, Avanzas Pablo, Angellotti Domenico, Esposito Giovanni, Vidal-Cales Pablo, Mohammadi Siamak, Farjat-Pasos Julio, Moris Cesar, Rodés-Cabau Josep. Optimal oversizing in transcatheter aortic valve replacement with the self-expanding Evolut valve system.. The Journal of invasive cardiology. 2024. DOI: 10.25270/jic/24.00011.