MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1155/2024/2172306

Медицинская библиотека · Cardiology research and practice · 2024

Performance and Physician Experience of INGEVITY+ Active Fixation Leads: Prospective INGEVITY+ Lead Clinical Study in Korea.

Joung Boyoung, Bae Myung Hwan, Oh Il-Young, Park Hyung-Seob, Shim Jaemin, Cho Min Soo, Lee Jung Myung, Choi Eue-Keun, Lee Young Soo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Boston Scientific INGEVITY+ pacing lead (Boston Scientific, Marlborough, MA, USA) has been upgraded to INGEVITY. The performance of the INGEVITY+ pacing lead has not yet been reported. This study aimed to evaluate the short- and long-term safety, effectiveness, and handling experience of INGEVITY+ leads.

METHODS

Consecutive patients were included from 9 institutions in Korea, where 400 leads (200 right ventricular active fixation leads and 200 right atrial active fixation leads) were implanted or attempted in 200 subjects.

RESULTS

During the implantation, only one patient required a lead change because of lead screw failure. The handling questionnaires of the lead received very positive feedback with 88% of operators agreeing that it is easy for leads to pass through small vessels or vessels with multiple leads. At the 3-month follow-up, 95.7% of RA leads and 99.5% of RV leads had pacing thresholds less than 1.5 V. A total of 92.4% of atrial leads had amplitudes greater than 1.5 mV, and 96.5% of ventricular leads had sensing amplitudes greater than 5 mV at 3 months. A total of 99.8% had impedances between 300 and 1,300 ohms. The lead-related complication-free rate for all leads during follow-up was 100%, and the overall rates of lead dislodgment, perforation, and pericardial effusion were all 0.0%.

CONCLUSIONS

The INGEVITY+ pacing lead exhibited exceptional clinical performance, with a high complication-free rate throughout the 3-month follow-up period. In addition, the lead displayed excellent electrical characteristics, and the lead-handling experience was reported to be very good.

Для цитирования: Joung Boyoung, Bae Myung Hwan, Oh Il-Young, Park Hyung-Seob, Shim Jaemin, Cho Min Soo, Lee Jung Myung, Choi Eue-Keun, Lee Young Soo. Performance and Physician Experience of INGEVITY+ Active Fixation Leads: Prospective INGEVITY+ Lead Clinical Study in Korea.. Cardiology research and practice. 2024. DOI: 10.1155/2024/2172306.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.