Safety verification of a novel irrigation catheter with flexible tip of laser-cut kerfs and contact force sensor.

Yamaguchi Junji, Takigawa Masateru, Goya Masahiko, Yamamoto Tasuku, Ikenouchi Takashi, Iwakawa Hidehiro, Negishi Miho, Goto Kentaro, Shigeta Takatoshi, Nishimura Takuro, Takamiya Tomomasa, Tao Susumu, Miyazaki Shinsuke, Sasano Tetsuo

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIMS: The safety evaluation of TactiFlex, a novel contact-force sensing catheter with a flexible 4-mm tip irrigated through laser-cut kerfs, has been ongoing. This study aimed to verify the safety of this type of catheter.

METHODS

Study 1: Radiofrequency (RF) applications at a range of powers (30-50 W), contact forces (10-20 g), and durations (10-60 s) using perpendicular/parallel catheter orientation with half-normal (HNS) or normal saline irrigation were compared between TactiFlex (4-mm tip) and TactiCath (3.5-mm tip) with temperature-controlled mode in excised porcine hearts. Study 2: The relation between RF applications using TactiFlex and the incidence of steam-pops in the real clinical cases were examined.

RESULTS

Study-1: 576 RF lesions were examined. TactiFlex demonstrated a significantly lower risk of steam-pops (5[1.7%] vs. 59[20.5%], p < .0001). Compared to 3.5-mm-tip catheter (TactiCath), 4-mm-tip catheter (TactiFlex) produced smaller lesion volume at perpendicular (193[98-554]mm3 vs. 263[139-436]mm3 , p < .0001), but relatively similar lesion volume at parallel contact (243[105-443]mm3 vs. 278[180-440]mm3 , p = .06). HNS-irrigation tended to increase the lesion volume in both catheters and to increase the incidence of steam-pops with TactiCath, but not with TactiFlex. The cut-off value of %impedance-drop ( = absolute impedance-drop/initial impedance) of 20% predicted steam-pops with a sensitivity = 100% and specificity = 89.6% in TactiFlex. Study-2: 5496 RF applications in 84 patients (51AFs/8ATs/3AVNRTs/4AVRTs/17PVCs/4VTs) using TactiFlex were analyzed. Four steam-pops (0.07%) in three patients with pericardial effusion were observed (%impedance-drop = 24%/26%/29%/35%, respectively). The cut-off value of %impedance-drop = 20%, derived from ex-vivo study, showed sensitivity = 100% and specificity = 90.1% in detecting steam-pops.

CONCLUSION

TactiFlex reduced the risk of steam-pops than TactiCath. %impedance-drop ≤ 20% may be reasonable for safely use with a sufficient safety margin. For 4-mm-tip catheter, parallel-contact may be recommended for larger lesion creation.

Для цитирования: Yamaguchi Junji, Takigawa Masateru, Goya Masahiko, Yamamoto Tasuku, Ikenouchi Takashi, Iwakawa Hidehiro, Negishi Miho, Goto Kentaro, Shigeta Takatoshi, Nishimura Takuro, Takamiya Tomomasa, Tao Susumu, Miyazaki Shinsuke, Sasano Tetsuo. Safety verification of a novel irrigation catheter with flexible tip of laser-cut kerfs and contact force sensor.. Pacing and clinical electrophysiology : PACE. 2023. DOI: 10.1111/pace.14868.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.