MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ajo.2016.04.004

Медицинская библиотека · American journal of ophthalmology · 2016

Central Corneal Thickness After Cross-linking Using High-Definition Optical Coherence Tomography, Ultrasound, and Dual Scheimpflug Tomography: A Comparative Study Over One Year.

Antonios Rafic, Fattah Maamoun Abdul, Maalouf Fadi, Abiad Bachir, Awwad Shady T

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess central corneal thickness (CCT) and corneal haze in keratoconus eyes following corneal cross-linking (CXL).

DESIGN

Prospective noncomparative case series.

METHODS

Forty patients (44 eyes) with keratoconus that had serial evaluation for CCT, following CXL, using high-definition optical coherence tomography (HD-OCT), ultrasound pachymetry (USP), and dual Scheimpflug tomography. CCT was measured at baseline, then at 2 weeks and 1, 3, 6, and 12 months postoperatively.

RESULTS

The mean baseline CCT measurements were 470.02 μm, 469.79 μm, and 466.66 μm using the HD-OCT, the USP, and dual Scheimpflug tomography, respectively (P = .91). Following CXL, the mean CCT measurements by dual Scheimpflug, at all follow-up periods, were lower compared to mean baseline reading (P < .003). The mean CCT measurements by the HD-OCT and USP were similar to baseline readings except for the thicker readings at the 2-week follow-up visit. Overall, the mean CCT obtained by the USP was similar to that obtained by the HD-OCT throughout the 12-month period. However, the mean CCT measurements obtained by dual Scheimpflug tomography were statistically and clinically significantly lower than those obtained by both HD-OCT and USP at all follow-up visits.

CONCLUSIONS

USP and HD-OCT showed better agreement in CCT readings at all visits, and may better estimate the true CCT following CXL as compared to dual Scheimpflug tomography. The lower mean CCT post-CXL as measured by the latter seems to be correlated with the amount of haze that develops after cross-linking.

Для цитирования: Antonios Rafic, Fattah Maamoun Abdul, Maalouf Fadi, Abiad Bachir, Awwad Shady T. Central Corneal Thickness After Cross-linking Using High-Definition Optical Coherence Tomography, Ultrasound, and Dual Scheimpflug Tomography: A Comparative Study Over One Year.. American journal of ophthalmology. 2016. DOI: 10.1016/j.ajo.2016.04.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.