Repeatability of Keratometry Measurements Obtained With Three Topographers in Keratoconic and Normal Corneas.

Guilbert Emmanuel, Saad Alain, Elluard Maud, Grise-Dulac Alice, Rouger Hélène, Gatinel Damien

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the repeatability of the corneal topography functions of Orbscan II (Bausch & Lomb, Rochester, NY), OPD-Scan III (Nidek, Gamagori, Japan), and iTrace (Tracey Technologies, Houston, TX) in keratoconic eyes and in a control group of normal patients.

METHODS

In this prospective cohort study, patients were recruited between November 2011 and May 2012. Measurements were performed with a combined Placido-scanning slit system (Orbscan II) and two combined Placido-aberrometer systems (OPD-Scan III and iTrace). Repeatability limit and intraclass correlation coefficients (ICCs) of keratometric readings were calculated.

RESULTS

Fifty-nine keratoconic eyes of 34 patients and 54 normal eyes of 27 patients were included. Three groups were evaluated: all stage I-IV keratoconic eyes (59 eyes), a subgroup consisting of stage I-II keratoconic eyes (41 eyes), and normal eyes (54 eyes). For almost all parameters studied, the repeatability limit was higher in the two groups of keratoconic eyes compared to normal eyes with all three topographers, indicating lower repeatability. For the maximum keratometry measurement, repeatability limit was 1.73, 1.49, and 1.41 diopters (D) in the stage I-IV keratoconic eyes group, 1.11, 1.02, and 0.98 D in the stage I-II keratoconic eyes group, and 0.61, 0.37, and 1.02 D in the normal eyes group with Orbscan II, OPD-Scan III, and iTrace, respectively.

CONCLUSIONS

Topographies performed in keratoconic eyes are less repeatable than those performed in normal eyes. Threshold values of keratometric changes used to ascertain keratoconus progression should be carefully considered. Caution should be taken when interpreting the topographies of such patients. The higher variability should be taken into account before performing any treatment.

Для цитирования: Guilbert Emmanuel, Saad Alain, Elluard Maud, Grise-Dulac Alice, Rouger Hélène, Gatinel Damien. Repeatability of Keratometry Measurements Obtained With Three Topographers in Keratoconic and Normal Corneas.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2016. DOI: 10.3928/1081597x-20160113-01.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.