DOI 10.1016/j.jcrs.2015.09.020
Accuracy of machine learning classifiers using bilateral data from a Scheimpflug camera for identifying eyes with preclinical signs of keratoconus.
Аннотация
PURPOSE: To describe the topographic and tomographic characteristics of normal fellow eyes of unilateral keratoconus cases and to evaluate the accuracy of machine learning classifiers in discriminating healthy corneas from the normal fellow corneas.
SETTING
Department of Ophthalmology, Semmelweis University, Budapest, Hungary.
DESIGN
Retrospective case-control study.
METHODS
Patients with bilateral keratoconus (keratoconus group), clinically and according to the keratoconus indices of the Pentacam HR Scheimpflug camera; normal fellow eyes of patients with unilateral keratoconus (fellow-eye group); and eyes of refractive surgery candidates (control group) were compared. Tomographic data, topographic data, and keratoconus indices were measured in both eyes using the Scheimpflug camera. Receiver operating characteristic (ROC) analysis was used to assess the performance of automated classifiers trained on bilateral data as well as individual parameters to discriminate fellow eyes of patients with keratoconus from control eyes.
RESULTS
Keratometry, elevation, and keratoconus indices values were significantly higher and pachymetry values were significantly lower in keratoconus eyes than in fellow eyes of unilateral keratoconus cases (P < .001). These fellow eyes had significantly higher keratometry, elevation, and keratoconus index values and significantly lower pachymetry values than control eyes (P < .001). Automated classifiers trained on bilateral data of index of height decentration had higher accuracy than the unilateral single parameter in discriminating fellow eyes of patients with keratoconus from control eyes (area under ROC 0.96 versus 0.88).
CONCLUSION
Automatic classifiers trained on bilateral data were better than single parameters in discriminating fellow eyes of patients with unilateral keratoconus with preclinical signs of keratoconus from normal eyes.
FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.