DOI 10.1016/j.ejcts.2008.05.004

Медицинская библиотека · European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2008

Do repeated runs of a cell saver device increase the pro-inflammatory properties of washed blood?

Vermeijden Wytze J, Hagenaars Ans, van Oeveren Willem, de Vries Adrianus J

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Intra-operative cell salvage is increasingly used, especially in longer cases with continuing blood loss. However it is unknown if the quality of processed blood is affected when larger quantities of blood are processed. We hypothesized that the quality of the washed blood decreases after multiple runs.

METHODS

Intra-operative cell salvage was performed in 42 consecutive patients undergoing cardiac surgery. When 1250 ml of blood was collected in the blood collection reservoir, this was processed and returned to the patient. In 21 patients more than 2500 ml of blood was collected during the whole procedure, thus allowing at least two subsequent runs with the auto-transfusion device. Blood samples were drawn from the blood collection reservoir of the cell saver device before, and from the processed blood after each run.

RESULTS

After the first run interleukin-6 concentrations were reduced with 85% (from 21+/-35 microg/l to 3.1+/-4.4 microg/l), whereas after the second run 72% was removed (63+/-69 microg/l to 17.6+/-25.3 microg/l). Leukocyte counts almost doubled after both processing runs (from 2.6+/-1.5 x 10(9)/l to 5+/-3.6 x 10(9)/l) and from 3.9+/-2.2 x 10(9)/l to 7.7+/-5.9 x 10(9)/l), hemoglobin concentration (14.8+/-1.6 mmol/l vs 15.0+/-1.1 mmol/l), free hemoglobin (2.3+/-1.6g/l vs 2.1+/-1.4 g/l) and platelet counts (18+/-9 x 10(9)/l vs 28+/-23 x 10(9)/l) were not different between the two runs.

CONCLUSIONS

Our results suggest, based on interleukin-6 and free hemoglobin washout that the quality of the processed blood remains constant with multiple runs of the cell saver device.

Для цитирования: Vermeijden Wytze J, Hagenaars Ans, van Oeveren Willem, de Vries Adrianus J. Do repeated runs of a cell saver device increase the pro-inflammatory properties of washed blood?. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2008. DOI: 10.1016/j.ejcts.2008.05.004.

Предметные рубрики MeSH