BACKGROUND

Monitoring of lead depth is crucial to achieve left bundle branch pacing (LBBP) with a low capture threshold and avoid septal perforation, but lacks informative approach.

OBJECTIVE

We aimed to prospectively assess the predictive value of current of injury on the occurrence of inadequate left bundle branch (LBB) capture threshold and acute septal perforation.

METHODS

Consecutive patients who received LBBP were enrolled. ST-segment elevation ≥ 25% of intrinsic R-wave amplitude on the unipolar intracardiac electrogram was defined as a sign of distinct current of injury. An LBB capture threshold of <1.5 V/0.5 ms was considered acceptable.

RESULTS

LBBP was attempted 513 times in 212 patients. The LBB capture threshold was more likely to improve to an acceptable level after 10 minutes in leads with initial (33 of 47 vs 0 of 8, with vs without) and residual (29 of 33 vs 4 of 14, with vs without) current of injury recorded on the tip electrode (P < .0001). Lead perforation during the procedure has occurred in 11 patients who had no current of injury noted on the tip electrode. The ratio of current of injury recorded on the tip electrode to that on the ring electrode was correlated to the lead depth determined by sheath angiography (Spearman correlation coefficient -0.624; P < .0001), and microperforation is highly possible when the ratio is decreased to <1 (sensitivity 100%; specificity 96.6%).

CONCLUSION

Current of injury is a useful tool in forecasting LBBP lead depth and septal perforation, and it could facilitate the decision-making process when the initial LBB capture threshold is undesirable.

Для цитирования: Shali Shalaimaiti, Wu Weiyun, Bai Jin, Wang Wei, Qin Shengmei, Wang Jingfeng, Liang Yixiu, Chen Haiyan, Su Yangang, Chen Xueying, Ge Junbo. Current of injury is an indicator of lead depth and performance during left bundle branch pacing lead implantation.. Heart rhythm. 2022. DOI: 10.1016/j.hrthm.2022.04.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.