Reframing Optimal Implantation of the Sapien 3 Transcatheter Heart Valve.
Аннотация
OBJECTIVES
To define the optimal implantation of the Sapien 3 (Edwards Lifesciences) transcatheter heart valve (THV), this study systematically analyzed the predeployment fluoroscopic THV position and correlated this to clinical outcomes.
METHODS
This was an observational study of 279 patients treated with the Sapien 3 THV. Fluoroscopic imaging was used to categorize patients into low (n = 147), intermediate (n = 86), and high (n = 46) implantation zones. These zones were based on the relationship of the balloon marker and radiolucent line of the valve frame (line of lucency) to the annular plane at deployment. The primary outcome was the rate of permanent pacemaker implantation (PPI) at 30 days. The secondary outcomes were the rates of new left bundle-branch block (LBBB) in-hospital and all-cause mortality at 1 year.
RESULTS
In the high, intermediate, and low groups, 30-day PPI rates were 4.3%, 8.1%, and 8.8% (P=.62); in-hospital LBBB rates were 10.9%, 26.7%, and 32.0% (P=.02); and all-cause mortality rates at 1 year were 3.1%, 7.3%, and 12.5% (P=.14), respectively. No differences were observed with respect to procedural success/complications or THV performance between the groups.
CONCLUSION
This study demonstrates fewer conduction abnormalities for Sapien 3 valves positioned within a higher zone defined fluoroscopically by the line of lucency and balloon marker.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.