MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euab267

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2022

Standard care vs. TRIVEntricular pacing in Heart Failure (STRIVE HF): a prospective multicentre randomized controlled trial of triventricular pacing vs. conventional biventricular pacing in patients with heart failure and intermediate QRS left bundle branch block.

Gould Justin, Claridge Simon, Jackson Thomas, Sieniewicz Benjamin J, Sidhu Baldeep S, Porter Bradley, Elliott Mark K, Mehta Vishal, Niederer Steven, Chadwick Humra, Kamdar Ravi, Adhya Shaumik, Patel Nikhil, Hamid Shoaib, Rogers Dominic, Nicolson William, Chan Cheuk F, Whinnett Zachary, Murgatroyd Francis, Lambiase Pier D, Rinaldi Christopher A

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

AIMS: To determine whether triventricular (TriV) pacing is feasible and improves CRT response compared to conventional biventricular (BiV) pacing in patients with left bundle branch block (LBBB) and intermediate QRS prolongation (120-150 ms).

METHODS AND

RESULTS

Between October 2015 and November 2019, 99 patients were recruited from 11 UK centres. Ninety-five patients were randomized 1:1 to receive TriV or BiV pacing systems. The primary endpoint was feasibility of TriV pacing. Secondary endpoints assessed symptomatic and remodelling response to CRT. Baseline characteristics were balanced between groups. In the TriV group, 43/46 (93.5%) patients underwent successful implantation vs. 47/49 (95.9%) in the BiV group. Feasibility of maintaining CRT at 6 months was similar in the TriV vs. BiV group (90.0% vs. 97.7%, P = 0.191). All-cause mortality was similar between TriV vs. BiV groups (4.3% vs. 8.2%, P = 0.678). There were no significant differences in echocardiographic LV volumes or clinical composite scores from baseline to 6-month follow-up between groups.

CONCLUSION

Implantation of two LV leads to deliver and maintain TriV pacing at 6 months is feasible without significant complications in the majority of patients. There was no evidence that TriV pacing improves CRT response or provides additional clinical benefit to patients with LBBB and intermediate QRS prolongation and cannot be recommended in this patient group.

CLINICAL TRIAL REGISTRATION NUMBER: Clinicaltrials.gov: NCT02529410.

Для цитирования: Gould Justin, Claridge Simon, Jackson Thomas, Sieniewicz Benjamin J, Sidhu Baldeep S, Porter Bradley, Elliott Mark K, Mehta Vishal, Niederer Steven, Chadwick Humra, Kamdar Ravi, Adhya Shaumik, Patel Nikhil, Hamid Shoaib, Rogers Dominic, Nicolson William, Chan Cheuk F, Whinnett Zachary, Murgatroyd Francis, Lambiase Pier D, Rinaldi Christopher A. Standard care vs. TRIVEntricular pacing in Heart Failure (STRIVE HF): a prospective multicentre randomized controlled trial of triventricular pacing vs. conventional biventricular pacing in patients with heart failure and intermediate QRS left bundle branch block.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2022. DOI: 10.1093/europace/euab267.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.