Intracardiac conduction time as a predictor of cardiac resynchronization therapy response: Results of the BIO|SELECT pilot study.

Soejima Kyoko, Kondo Yusuke, Sasaki Shingo, Adachi Kazumasa, Kato Ritsushi, Hagiwara Nobuhisa, Harada Tomoo, Kusano Kengo, Miura Fumiharu, Morishima Itsuro, Yoshitani Kazuyasu, Yotsukura Akihiko, Fujimoto Manabu, Nishii Nobuhiro, Shimeno Kenji, Ohe Masatsugu, Tasaka Hiroshi, Sasaki Hiroto, Schrader Juergen, Ando Kenji

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Quadripolar left ventricular (LV) leads are capable of sensing and pacing the left ventricle from 4 different electrodes, which may potentially improve patient response to cardiac resynchronization therapy (CRT).

OBJECTIVE

We measured 3 different time intervals: right ventricular (RV)-sensed to LV-sensed during intrinsic rhythm (RVs-LVs), RV-paced to LV-sensed (RVp-LVs), and LV-paced to LV-sensed (LVp-LVs, between distal [LV1] and proximal pole on a quadripolar LV lead), and assessed their association with CRT response in terms of LV end-systolic volume (LVESV) and a composite benefit index (CBI) comprising LVESV, LV ejection fraction (LVEF), brain natriuretic peptide level, and NYHA class.

METHODS

A CRT-defibrillator system with quadripolar LV lead was implanted in 196 patients (mean age 69 years, mean LVEF 30%, left bundle-branch block [LBBB] 58%). Conduction intervals were measured before hospital discharge. At baseline and 7-month follow-up, echocardiographic and other components of CBI were determined.

RESULTS

The mean RVs-LV1s, RVp-LV1s, and LVp-LVs delays were 68 ± 38 ms, 132 ± 34 ms, and 99 ± 31 ms, respectively. From baseline to 7 months, LVESV decreased by 17.3% ± 28.6%. The RVs-LV1s interval correlated stronger with CBI (R2 = 0.12, P < .00001) than with LVESV change (R2 = 0.05, P = .006). In contrast, RVp-LV1s did not correlate and LVp-LVs correlated only weakly with CRT response. The subgroup of patients (44%) with LBBB and RVs-LV1s above the lower quartile (≥34 ms) showed the greatest response to CRT.

CONCLUSION

The RVs-LVs interval during intrinsic rhythm is relevant for CRT success, whereas RVp-LVs and LVp-LVs intervals did not predict CRT response.

Для цитирования: Soejima Kyoko, Kondo Yusuke, Sasaki Shingo, Adachi Kazumasa, Kato Ritsushi, Hagiwara Nobuhisa, Harada Tomoo, Kusano Kengo, Miura Fumiharu, Morishima Itsuro, Yoshitani Kazuyasu, Yotsukura Akihiko, Fujimoto Manabu, Nishii Nobuhiro, Shimeno Kenji, Ohe Masatsugu, Tasaka Hiroshi, Sasaki Hiroto, Schrader Juergen, Ando Kenji. Intracardiac conduction time as a predictor of cardiac resynchronization therapy response: Results of the BIO|SELECT pilot study.. Heart rhythm O2. 2021. DOI: 10.1016/j.hroo.2021.09.007.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.