MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.oftal.2014.09.007

Медицинская библиотека · Archivos de la Sociedad Espanola de Oftalmologia · 2014

[Correlating cup-to-disc ratios measured by HRT-III, SD-OCT and the new color imaging Laguna ONhE procedure].

Rodríguez Uña I, Méndez Hernández C D, Sáenz-Francés F, García Feijóo J

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To examine correlations between cup-to-disc (C/D) ratios determined by the new Laguna ONhE (optic nerve hemoglobin) color imaging procedure, spectral domain optical coherence tomography (OCT), confocal scanning laser tomography using Heidelberg retina tomography (HRT), and examining retinal images.

METHODS

C/D ratio measurements were made on 154 eyes of 154 subjects (52 healthy controls, 36 with ocular hypertension and 66 with primary open-angle glaucoma) using the Laguna ONhE, HRT-III (Heidelberg Engineering) and OCT Spectralis (Heidelberg Engineering) instruments and photographs of the optic disc were examined by a blinded observer (experienced glaucoma specialist).

RESULTS

Global intraclass correlation coefficients (ICC) were: 0.379 (95% CI: 0.233-0.508) for Laguna ONhE-HRT, 0.621 (95% CI: 0.513-0.709) for Laguna ONhE-OCT, and 0.558 (95% CI: 0.398-0.678) for the Laguna ONhE-observer, indicating significant agreement in each case (P<.001). The highest ICC was recorded for OCT- observer (0.715; 95% CI: 0.605-0.795).

CONCLUSIONS

C/D ratios measured using the Laguna ONhE procedure correlated well with OCT measurements and retinography measurements made by an experienced observer. Best correlation was observed for OCT versus observer measurements. Agreement was good between the Laguna ONhE, OCT and observer measurements, and was somewhat lower between HRT and the remaining procedures.

Для цитирования: Rodríguez Uña I, Méndez Hernández C D, Sáenz-Francés F, García Feijóo J. [Correlating cup-to-disc ratios measured by HRT-III, SD-OCT and the new color imaging Laguna ONhE procedure].. Archivos de la Sociedad Espanola de Oftalmologia. 2014. DOI: 10.1016/j.oftal.2014.09.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.