[Investigation of Metal Artifacts Occurring in Axial Planes without Intracranial Metal Coils: A Phantom Study Using Area Detector CT].
Аннотация
PURPOSE: The objective of this study was to evaluate metal artifacts on axial planes without visible metal by simulating postoperative computed tomography (CT) examinations following coil embolization of dural arteriovenous fistulas. Specifically, we assessed the behavior of metal artifacts in the pericoil region and aimed to propose an imaging protocol that minimizes metal artifact occurrence.
METHODS
An intracranial metallic coil phantom was constructed and scanned using two different area detector CT systems. The behavior of metal artifacts around the coil was evaluated according to the axial position of the metal coil, collimation width, reconstruction algorithm, scan mode, and the presence or absence of metal artifact reduction (MAR).
RESULTS
We confirmed that CT value elevation occurred on axial planes without visible metal due to artifacts originating from the metallic coil. This phenomenon was observed regardless of coil position, scan mode, or the use of MAR. It also occurred with all reconstruction algorithms except model-based iterative reconstruction. Furthermore, narrowing the collimation width reduced the magnitude of CT value elevation.
CONCLUSION
Because artifacts occurring on axial planes without visible metal may be reduced by using a narrower collimation width, scanning with narrow collimation is recommended for postoperative CT examinations following cerebral aneurysm embolization.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.