COMBINED APPLICATION OF B-SCAN ULTRASONOGRAPHY AND EYE-STEERING ULTRAWIDE FIELD IMAGING TO IMPROVE THE DETECTION OF RETINAL TEARS BEFORE CATARACT SURGERY.

Meng Jiaqi, Cheng Kaiwen, Huang Zhiqian, He Wenwen, Zhang Keke, Lu Yi, Zhu Xiangjia

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the efficacy of combined application of B-scan ultrasonography (US) and ultrawide field imaging (UWFI) in detecting retinal tears before cataract surgery.

METHODS

Of 1,277 cataract patients, 2,552 eyes were enrolled and received both B-scan US and UWFI examinations preoperatively. Three types of combination were applied: type 1 (union, B-scan US or centered UWFI), type 2 (intersection, B-scan US and centered UWFI), and type 3 (B-scan US and eye-steering UWFI). Sensitivity and specificity of detecting retinal tears by different methods were assessed.

RESULTS

Totally 4.55% (116/2,552) of eyes were presented with retinal tears. The sensitivity of B-scan US and UWFI was 87.93% and 84.48%, and specificity was 95.16% and 99.79%, respectively. By applying type 1 and type 2 combination, the sensitivity was 98.28% and 74.14%, and specificity was 95.03% and 99.92%, respectively. By type 3 combination, the sensitivity increased to 95.69% and specificity to 99.88%, both of which were comparable to indirect ophthalmoscopy regardless of the number, type, and location of tears ( P > 0.05). In eyes with any cataract type or axial length, type 3 combination also gained comparable performance to indirect ophthalmoscopy.

CONCLUSION

Combined application of B-scan US and eye-steering UWFI presented satisfactory performance in detecting retinal tears before cataract surgery.

Для цитирования: Meng Jiaqi, Cheng Kaiwen, Huang Zhiqian, He Wenwen, Zhang Keke, Lu Yi, Zhu Xiangjia. COMBINED APPLICATION OF B-SCAN ULTRASONOGRAPHY AND EYE-STEERING ULTRAWIDE FIELD IMAGING TO IMPROVE THE DETECTION OF RETINAL TEARS BEFORE CATARACT SURGERY.. Retina (Philadelphia, Pa.). 2024. DOI: 10.1097/iae.0000000000004040.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.