Robotic system for nasopharyngeal swab sampling based on remote center of motion mechanism.

Shim Seongbo, Seo Joonho

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: In this study, a robotic system is proposed for nasopharyngeal (NP) swab sampling with high safety and efficiency. Most existing swab-sampling robots have more than six degrees of freedom (DOFs). However, not all six DOFs are necessarily required for NP swab sampling. A high number of DOFs can cause safety problems, such as collisions between the robot and patient.

METHOD: We developed a new type of robot with four DOFs for NP swab sampling that consists of a two DOFs remote center of motion (RCM) mechanism, a two DOFs insertion mechanism, and a nostril support unit. With the nostril support unit, the robot no longer needs to adjust the insertion position of the swab. The proposed robot enables the insertion orientation and depth to be adjusted according to different postures or facial shapes of the subject. For intuitive and precise remote control of the robot, a dedicated master device for the RCM and a visual feedback system were developed.

RESULT: The effectiveness of the robotic system was demonstrated by repeatability, RCM accuracy, tracking accuracy, and in vitro phantom experiments. The average tracking error between the master device and the robot was less than 2 mm. The contact force exerted on the swab prior to reaching the nasopharynx was less than 0.04 N, irrespective of the phantom's pose.

CONCLUSION

This study confirmed that the RCM-based robotic system is effective and safe for NP swab sampling while using minimal DOFs.

Для цитирования: Shim Seongbo, Seo Joonho. Robotic system for nasopharyngeal swab sampling based on remote center of motion mechanism.. International journal of computer assisted radiology and surgery. 2023. DOI: 10.1007/s11548-023-03032-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.