Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry.

Gao Rongrong, Chen Hao, Savini Giacomo, Miao Yaxin, Wang Xiaorui, Yang Jing, Zhao Weiqi, Wang Qinmei, Huang Jinhai

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

The purpose of the current study was to compare the measurements between a new optical biometer based on swept-source optical coherence tomography (SS-OCT), the OA-2000 (Tomey, Japan), and an optical biometer based on optical low coherence reflectometry (OLCR), the Lenstar (Haag-Streit, Switzerland). Ninety-nine eyes of 99 healthy subjects were included. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), keratometry (K) readings, including flat K (Kf), steep K (Ks), mean K (Km), astigmatism vectors J0, J45 at diameters of 2.5 and 3.0 mm, and white-to-white diameter (WTW) were measured three times each using both biometer in normal eyes by random sequence. Bland-Altman analysis showed good agreement between the SS-OCT and OLCR devices for AL, AD, ACD, LT, with narrow 95% LoA (-0.05 to 0.07 mm, -0.09 to 0.10 mm, -0.10 to 0.09 mm, and -0.06 to 0.22 mm, respectively), and the P values of ACD were both >0.05. The CCT, Kf, Ks, Km, J0, J45 and WTW values provided by the OA-2000 were in good agreement with the Lenstar, and statistically significant differences were detected for some of them but not clinical differences. The agreement was excellent especially for AL.

Для цитирования: Gao Rongrong, Chen Hao, Savini Giacomo, Miao Yaxin, Wang Xiaorui, Yang Jing, Zhao Weiqi, Wang Qinmei, Huang Jinhai. Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry.. Scientific reports. 2017. DOI: 10.1038/s41598-017-02463-z.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.