Posterior corneal shape: Comparison of height data from 3 corneal topographers.
Аннотация
PURPOSE: To compare the ability of 3 clinical corneal topographers to describe the posterior corneal shape.
SETTING
University Medical Center Groningen, the Netherlands.
DESIGN
Prospective observational study.
METHODS
Corneas of healthy participants were measured twice with a dual Scheimpflug instrument (Galilei G2), a scanning-slit system (Orbscan IIz), and a single Scheimpflug instrument (Pentacam HR). Height data describing the posterior corneal shape were fit with Zernike polynomials. Mean values with standard deviations (SD), test-retest variability (coefficient of repeatability [CoR]), and interdevice variability were determined for the defocus Z(2,0), astigmatism Z(2,-2) and Z(2,2), and higher-order terms coma Z(3,-1) and Z(3,1), trefoil Z(3,-3) and Z(3,3), and spherical aberration Z(4,0) coefficients for 5.5 mm and 8.0 mm diameters.
RESULTS
For the 5.5 mm diameter, CoRs ranged from 0.3 to 4.3 μm with the dual Scheimpflug instrument, 1.6 to 5.2 μm with the scanning-slit system, and 0.3 to 2.0 μm with the single Scheimpflug instrument. The CoR was similar for the Scheimpflug instruments (P = .43) but poorer for the scanning-slit system (P < .001). The CoRs of the Scheimpflug instruments were smaller than the corresponding population SD for defocus, cardinal astigmatism, coma, and spherical aberration. The scanning-slit system failed to provide 8.0 mm diameter data. There was a significant bias (interdevice variability) between the Scheimpflug instruments in the higher-order coefficients at both diameters.
CONCLUSIONS
Repeatability in assessing the posterior corneal shape was generally good for the Scheimpflug instruments but poor for the scanning-slit system. Interdevice variability between the Scheimpflug instruments compromised the interchangeability of higher-order coefficients. For astigmatism, CoR and 95% limits of agreement of the Scheimpflug instruments typically corresponded to 0.1 diopter per astigmatism term.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.