Corneal flap thickness with the Moria M2 single-use head 90 microkeratome in 72 consecutive LASIK procedures.
Аннотация
PURPOSE: This study aimed to evaluate the accuracy and consistency of corneal flap thickness in laser-assisted in situ keratomileusis (LASIK) with the Moria M2 single-use head 90 microkeratome.
METHODS
The central corneal thickness of 72 (37 right, 35 left) eyes of 37 patients was measured by ultrasonic pachymetry preoperatively and intraoperatively after flap cut. The Moria M2 single-use head 90 microkeratome was used to create a superior hinged flap in all eyes. The right eyes were always operated on before the left eyes in each patient, using the same blade in all bilateral cases. All patients underwent LASIK for myopia and/or myopic astigmatism using VISX Star S4 platform.
RESULTS
The mean preoperative spherical equivalent refraction was -3.55±2.30 D (range: -0.625 to -11.00 D), preoperative central corneal thickness by ultrasonic pachymetry was 541±26.82 µm (490-600 µm) and steepest K was 44.08±1.49 D (40-46.75 D) in all eyes. The mean flap thickness was 136.97±20.07 µm (106-192 µm), 131.2±19.5 µm (91-192 µm), and 134.16±19.85 µm (91-192 µm) in the right, left, and both eyes, respectively. A positive significant relationship was found between flap thickness and preoperative ultrasonic pachymetry thickness. No significant relationship was found between flap thickness and the age, preoperative spherical equivalent, and preoperative steepest K. The difference between the first and second eyes was not significant. There were no major intraoperative and postoperative complications in all eyes.
CONCLUSION
The Moria M2 single-use head 90 microkeratome cut relatively thicker flaps than were intended. The flap thickness range was quite wide. This was a disadvantage for the accuracy and consistency of corneal flap thickness.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.