DOI 10.1016/j.jcrs.2016.12.011

Медицинская библиотека · Journal of cataract and refractive surgery · 2017

Rotational stability of 2 intraocular lenses with an identical design and different materials.

Draschl Petra, Hirnschall Nino, Luft Nikolaus, Schuschitz Sandra, Wiesinger Jörg, Rigal Karl, Findl Oliver

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the rotational stability of nontoric intraocular lenses (IOLs) of the same design and different materials.

SETTING

Vienna Institute for Research in Ocular Surgery, Department of Ophthalmology, Hanusch Hospital, Vienna, Austria.

DESIGN

Prospective randomized case series.

METHODS

This study included cataract patients with a corneal astigmatism of less than 1.75 diopters measured with the IOLMaster 500. Each patient received a hydrophilic IOL (Pod Ay 26P) in 1 eye and a hydrophobic IOL (Podeye) in the other eye. One hour and 3 months postoperatively, retroillumination photographs were taken to assess rotational stability of the IOLs. In addition, autorefraction, subjective refraction, and Purkinje meter measurements were performed at the 3-month follow-up.

RESULTS

Eighty eyes of 40 patients were included in this study. Three months postoperatively, the IOL rotation within the first 3 months was 2.4 degrees ± 1.85 (SD) in the hydrophilic IOL group and 1.6 ± 1.61 degrees in the hydrophobic IOL group. This difference was statistically significant (P = .016).

CONCLUSION

The hydrophobic IOL was rotationally more stable than the hydrophilic IOL, although both IOLs provided good capsular bag stability.

Для цитирования: Draschl Petra, Hirnschall Nino, Luft Nikolaus, Schuschitz Sandra, Wiesinger Jörg, Rigal Karl, Findl Oliver. Rotational stability of 2 intraocular lenses with an identical design and different materials.. Journal of cataract and refractive surgery. 2017. DOI: 10.1016/j.jcrs.2016.12.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.