Effect of storage temperatures simulating transport conditions of nasopharyngeal swabs on the results of a chemiluminescence immunoassay (CLIA) to detect severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) antigen.

Kanta Poonam, Ghosh Arnab, Goyal Kapil, Kaur Ramandeep, Baluni Manjari, Sircar Shubhankar, Kumar Vikas, Kumar Krishan, Vijay Neetu, Singh Mini P

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Chemiluminescence Immunoassay (CLIA) is high throughput, rapid diagnostic test which has recently come up for the detection of SARS-CoV-2 antigen. The present study evaluated performance of CLIA antigen test in nasopharyngeal swab samples stored at different temperatures for 7 days to simulate the transport conditions and transit time across the country from remote peripheral laboratories to central facilities.

MATERIALS AND

METHODS

Limit of detection (LOD), sensitivity and specificity of VITROS® SARS-CoV-2 antigen assay was determined using Real-time reverse transcriptase PCR (rRT-PCR) confirmed SARS-CoV-2 positive and negative samples. To detect the effect of storage temperatures on VITROS ®SARS-CoV-2 antigen results, samples were stored at 4 ​°C, 25 ​°C & 37 ​°C for 7 days followed by detection of SARS-CoV-2 nucleocapsid antigen and compared with N-gene rRT-PCR.

RESULTS

The VITROS® SARS-CoV-2 antigen test was found to have a sensitivity and specificity of 78.9% and 100% respectively with high sensitivity of 88.1% for samples with Ct ​< ​30. The LOD of VITROS assay was equivalent to 3800 copies of RNA per reactions as compared to 72 copies per reaction for rRT-PCR. We observed that more than 80% of samples with <30 Ct values could be detected by VITROS SARS-CoV-2 antigen assay at day 7 even when stored at 37 ​°C. For samples with Ct values between 26 and 30, on day 7 the positivity rate of N-antigen at 4 ​°C was 90.9% and 37 ​°C was 63.6%.

CONCLUSIONS

CLIA testing can be carried out for the detection of SARS-CoV-2 N-protein in NP-swab samples transported in cold chain even with 7 days transit time, particularly for Ct ​< ​30 samples which represents cases with higher transmissibility. As drop in positivity for VITROS assay was lower as compared to rRT-PCR on day 7 in cold chain-maintained samples, the assay can be useful to screen samples received from remote peripheral areas before performing rRT-PCR.

Для цитирования: Kanta Poonam, Ghosh Arnab, Goyal Kapil, Kaur Ramandeep, Baluni Manjari, Sircar Shubhankar, Kumar Vikas, Kumar Krishan, Vijay Neetu, Singh Mini P. Effect of storage temperatures simulating transport conditions of nasopharyngeal swabs on the results of a chemiluminescence immunoassay (CLIA) to detect severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) antigen.. Indian journal of medical microbiology. 2023. DOI: 10.1016/j.ijmmb.2023.100379.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.