Comparison of Axial Length, Corneal Curvature, and Anterior Chamber Depth Measurements of 2 Recently Introduced Devices to a Known Biometer.

Shajari Mehdi, Cremonese Carla, Petermann Kerstin, Singh Pankaj, Müller Michael, Kohnen Thomas

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: Evaluate agreement and repeatability of axial length (AL), corneal curvature, and anterior chamber depth measurements of 2 new devices, 1 using Scheimpflug technology combined with partial coherence interferometry (Pentacam AXL) and 1 using swept-source optical coherence tomography (IOL Master 700) to a widely used optical biometer (IOL Master 500).

DESIGN

Prospective, instrument reliability analysis.

METHODS

This study was conducted in the Department of Ophthalmology at Goethe-University, Frankfurt, Germany. Seventy-nine eyes in 79 patients were measured with the IOL Master 500, the IOL Master 700, and with the Pentacam AXL. All eyes were measured twice with each device. The results were statistically compared using Bland-Altman plots. Outcome measures included AL, corneal curvature, and the measurement of anterior chamber depth.

RESULTS

The Bland-Altman analysis of axial length measurements with the IOL Master 500 and IOL Master 700 showed a mean difference of 0.008 mm between the 2 devices, with a coefficient of repeatability (COR) of 0.044 (P = .05). For measurements with the IOL Master 500 and Pentacam AXL, a mean difference of 0.026 mm and COR of 0.055 (P = .75) was found. For measurements with Pentacam AXL and IOL Master 700, a mean difference of -0.019 mm and COR of 0.052 (P = .06) was found. In terms of repeatability, there was no significant difference in measurement of any variable with any device. Vector analysis showed no significant difference for corneal astigmatism measurements (P = .84).

CONCLUSION

There was no statistically significant difference in compared biometric parameters performed with all 3 devices.

Для цитирования: Shajari Mehdi, Cremonese Carla, Petermann Kerstin, Singh Pankaj, Müller Michael, Kohnen Thomas. Comparison of Axial Length, Corneal Curvature, and Anterior Chamber Depth Measurements of 2 Recently Introduced Devices to a Known Biometer.. American journal of ophthalmology. 2017. DOI: 10.1016/j.ajo.2017.02.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.