DOI 10.1016/j.resuscitation.2004.04.005

Медицинская библиотека · Resuscitation · 2004

Preliminary in-hospital experience with a fully automatic external cardioverter-defibrillator.

Bento André Moreira, Cardoso Luiz Francisco, Timerman Sérgio, Moretti Miguel Antonio, Peres Eduardo Dante Bariani, de Paiva Edison Ferreira, Ramires José Antonio Franchini, Kern Karl B

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Ventricular fibrillation (VF) and ventricular tachycardia (VT) are frequently present as initial rhythms during in-hospital cardiac arrest. Although ample evidence exists to support the need for rapid defibrillation, the response to in-hospital cardiac arrest remains without major advances in recent years. The delay between the arrhythmic event and intervention is still a challenge for clinical practice.

OBJECTIVE

To analyze the performance and safety of in-hospital use of a programmable, fully automatic external cardioverter-defibrillator (AECD).

METHODS

We conducted a prospective study at the Emergency Department of a university hospital. A total of 55 patients considered to be at risk of sustained VT/VF were included. Patients underwent monitoring of their cardiac rhythm by the AECD. Upon detection of a ventricular tachyarrhythmia, the AECD was programmed to automatically deliver shock therapy.

RESULTS

We recorded 19 episodes of VT/VF in 3 patients. The median time between the beginning of the arrhythmia and the first defibrillation was 33.4 s (21-65 s). One episode of spontaneous reversion of VT was documented 20 s after its origin and shock therapy was aborted. The defibrillation success was 94.4% (17/18) for the first shock and 100% (1/1) for the second shock. No case of inappropriate shock discharge was registered during the study period.

CONCLUSION

The AECD has the feasibility to combine long-term monitoring with automatic defibrillation safely and effectively. It presents the possibility of providing rapid identification of, and response to, in-hospital ventricular tachyarrhythmias.

Для цитирования: Bento André Moreira, Cardoso Luiz Francisco, Timerman Sérgio, Moretti Miguel Antonio, Peres Eduardo Dante Bariani, de Paiva Edison Ferreira, Ramires José Antonio Franchini, Kern Karl B. Preliminary in-hospital experience with a fully automatic external cardioverter-defibrillator.. Resuscitation. 2004. DOI: 10.1016/j.resuscitation.2004.04.005.