Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications.

Ellenbogen Kenneth A, Wood Mark A, Shepard Richard K, Clemo Henry F, Vaughn Tim, Holloman Keith, Dow Marlene, Leffler Jeanne, Abeyratne Athula, Verness Dave

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study evaluated the long-term reliability of an implantable cardioverter defibrillator (ICD) lead to determine the incidence, clinical presentation, and management of lead failure.

BACKGROUND

Despite recent advances in ICD technology, the long-term reliability of ICD leads remains a significant problem.

METHODS

Concern about long-term reliability of coaxial polyurethane ICD leads caused us to systematically study all patients implanted with Medtronic (Minneapolis, Minnesota) 6936 lead at our institution. We performed follow-up of 74 patients with 76 ICD leads that were implanted from February 28, 1995 to September 8, 1997. Thirty-seven patients underwent routine clinical ICD follow-up testing and ventricular fibrillation induction to determine the status of their ICD lead after a mean follow-up of 68.6 +/- 8.2 months.

RESULTS

The lead survival analysis shows a cumulative failure probability of 37% (confidence interval, 24% to 54%) at 68.6 months. Six patients demonstrated a previously undescribed mode of ICD lead failure: prolonged oversensing immediately after shock therapy. The use of short interval counters to monitor nonphysiologic R-R intervals and the measurement of ring-to-coil impedance detected early lead failures in five patients.

CONCLUSIONS

This analysis shows: 1) problems with ICD leads may not become apparent until late during follow-up and may become a significant late problem, 2) a "signature" mode of lead failure for the 6936 consisting of oversensing of electrical noise following shocks, 3) early detection of lead failure with a short interval counter algorithm or measurement of ring-to-coil impedance may be clinically useful.

Для цитирования: Ellenbogen Kenneth A, Wood Mark A, Shepard Richard K, Clemo Henry F, Vaughn Tim, Holloman Keith, Dow Marlene, Leffler Jeanne, Abeyratne Athula, Verness Dave. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications.. Journal of the American College of Cardiology. 2003. DOI: 10.1016/s0735-1097(02)02620-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.