Clinical experience with a dual-chamber implantable cardioverter defibrillator to treat atrial tachyarrhythmias.
Аннотация
INTRODUCTION
This study evaluated the safety and efficacy of a new dual-chamber implantable cardioverter defibrillator (ICD) to detect and treat atrial tachyarrhythmias in patients with drug-refractory atrial fibrillation (AF) and no indication for a ventricular ICD.
METHODS AND
RESULTS
A dual-chamber ICD (Medtronic 7250 Jewel AF) was implanted in 144 of 146 patients. The device discriminates atrial tachycardia from AF based on cycle length and regularity, and uses atrial overdrive pacing as well as shocks to terminate tachyarrhythmia episodes. Patients were followed for an average of 12.6+/-6.2 months. Use of antiarrhythmic drugs was 63% at baseline and did not change over time. Kaplan-Meier estimates of 12-month complication-free survival, device therapy survival, and patient survival were 85%, 91%, and 98%, respectively. Positive predictive accuracy of spontaneous atrial tachyarrhythmia detection was 99%, while atrial overdrive pacing and shocks terminated 40% and 87% of treated episodes, respectively. Median duration of successfully treated episodes was 8.9 minutes versus 144 minutes for the therapy failures. There was no reduction in the use of patient-activated shock therapy over time; at the 12-month follow-up evaluation, 94% of patients were in sinus rhythm. Ventricular tachyarrhythmias (67 episodes) were detected and appropriately treated in 7.6% of patients.
CONCLUSION
This dual-chamber ICD appears to be safe and well tolerated in patients with drug-refractory symptomatic atrial tachyarrhythmias. The device, used in combination with drugs, effectively treats atrial tachyarrhythmias with pacing and/or shock therapies and decreases the median episode duration. In addition, the device protects from ventricular tachyarrhythmias in patients with AF and structural heart disease.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.