DOI 10.1007/s11325-011-0558-1

Медицинская библиотека · Sleep & breathing = Schlaf & Atmung · 2011

Multi-modal ECG Holter system for sleep-disordered breathing screening: a validation study.

Poupard Laurent, Mathieu Marc, Goldman Michael, Chouchou Florian, Roche Frédéric

Сравнительное исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The high prevalence of sleep disordered breathing (SDB) among heart diseases patients becomes increasingly recognized. A reliable exploring tool of SDB well adapted to cardiologists practice would be very useful for the management of these patients.

METHODS

We assessed a novel multi-modal electrocardiogram (ECG) Holter which incorporated both thoracic impedance and pulse oximetry signals. We compared in a home setting, a standard condition for Holter recordings, results from the novel device to a classical ambulatory polygraph in subjects with suspected SDB. The analysis of cardiac arrhythmias in relationship with SDB is also presented. A total of 118 patients clinically suspected of having SDB were evaluated (mean age 57 ± 14 years, mean body mass index [BMI] 32 ± 6 kg/m(2)). The new device allows calculating a new index called thoracic impedance (TI) disturbance index (TIDI+) evaluated from TI and SpO(2) signals recorded from a Holter monitor.

RESULTS

In the population under study, 93% had more than 70% of usable TI signal and 95% had more than 90% for SpO(2) during sleep time recording. Screening performance results based on automatic analysis is accurate: TIDI + demonstrates a high level of sensitivity (96.8%), specificity (72.3%) as well as positive (82.4%) and negative (94.4%) predictive value for the detection of SDB. Moreover, detection of SDB periods permits us to observe a possible respiratory association of several nocturnal arrhythmias.

CONCLUSIONS

The multi-modal Holter should be considered as a valuable evaluating tool for SDB screening and as a case selection technique for facilitating access to a full polysomnography for severe cases. Moreover, it offers a unique opportunity to study arrhythmia consequences with both respiratory and hypoxia disturbances.

Для цитирования: Poupard Laurent, Mathieu Marc, Goldman Michael, Chouchou Florian, Roche Frédéric. Multi-modal ECG Holter system for sleep-disordered breathing screening: a validation study.. Sleep & breathing = Schlaf & Atmung. 2011. DOI: 10.1007/s11325-011-0558-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.