Evaluation of stability and potential interference on the α-thalassaemia early eluting peak and immunochromatographic strip test for α-thalassaemia --SEA carrier screening.
СтатьяОценочное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lam Wing Kit, Fan Christina Pui Ying, Law Winnie Yim Fong, Wong Tsz Fung, Too Darcy Lok Han, Ko Lok Nga, Leung Anskar Yu Hung, Yip Sze Fai
Аннотация
OBJECTIVES: α-Thalassaemia screening is crucial for identifying carriers at risk of having offspring with haemoglobin (Hb) Bart's hydrops fetalis syndrome. This study evaluated the performance of two potential screening methods: the α-thalassaemia early eluting peak (αEEP) identified by high-performance liquid chromatography (HPLC) and an immunochromatographic strip test (ICT), focusing on stability and potential interferences.
METHODS: Ninety-two peripheral blood samples were used. Thirty were for assessment of αEEP and ICT stability and interference by icterus/lipaemia, and saline washing for interference removal. Diagnostic performance of αEEP and ICT were evaluated in 40 samples with glycated Hb (P2) ≥6.0 % and 22 with Hb F≥2.0 % on HPLC, using α-globin genotyping as the gold standard.
RESULTS: Both αEEP and ICT results remained stable for 14 days. Neither test was significantly affected by icterus or lipaemia, though 3 of 20 ICT results (15 %) showed discordance after saline washing. Elevated HbA1c and Hb F altered early eluting peak patterns but did not affect αEEP interpretation. For detecting --SEA mutation, αEEP showed 100 % sensitivity and 100 % specificity when P2≥6.0 % or Hb F≥2.0 %, while ICT a low specificity (45 %) when Hb F≥2.0 %.
CONCLUSIONS: αEEP showed reliable performance in detecting α0-thalassaemia with --SEA mutation in all conditions tested, while ICT showed low specificity when Hb F≥2.0 %. These findings support αEEP as a reliable test for routine clinical laboratory use, while cautions should be made for ICT in case of elevated Hb F levels.
METHODS: Ninety-two peripheral blood samples were used. Thirty were for assessment of αEEP and ICT stability and interference by icterus/lipaemia, and saline washing for interference removal. Diagnostic performance of αEEP and ICT were evaluated in 40 samples with glycated Hb (P2) ≥6.0 % and 22 with Hb F≥2.0 % on HPLC, using α-globin genotyping as the gold standard.
RESULTS: Both αEEP and ICT results remained stable for 14 days. Neither test was significantly affected by icterus or lipaemia, though 3 of 20 ICT results (15 %) showed discordance after saline washing. Elevated HbA1c and Hb F altered early eluting peak patterns but did not affect αEEP interpretation. For detecting --SEA mutation, αEEP showed 100 % sensitivity and 100 % specificity when P2≥6.0 % or Hb F≥2.0 %, while ICT a low specificity (45 %) when Hb F≥2.0 %.
CONCLUSIONS: αEEP showed reliable performance in detecting α0-thalassaemia with --SEA mutation in all conditions tested, while ICT showed low specificity when Hb F≥2.0 %. These findings support αEEP as a reliable test for routine clinical laboratory use, while cautions should be made for ICT in case of elevated Hb F levels.