Assessment of fundal accommodation in solid meal gastric emptying scintigraphy: a retrospective analysis.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Farghaly Hussein, Alahmari Abdullah, Albatly Abdulrahman, Almukhailed Omar, Almutairy Sami, Mohamadien Nsreen R A, Nasr Hatem
Аннотация
AIM: To investigate the feasibility of both quantitative and qualitative assessment of fundal accommodation during solid meal gastric emptying (SGE) scintigraphy.
METHODS: SGE scintigraphy studies of 72 patients were retrospectively evaluated. Regions of interest (ROIs) were generated on the stomach at different times (0, 30 min, 1, 2, 3, and 4 h), and the corresponding retention percentages were calculated. Counts as geometric means from anterior and posterior ROIs over proximal stomach half and whole stomach at zero-time were used to calculate gastric fundal accommodation ratio (FAR). The radioactivity at the fundus was visually assessed at zero-time image to generate a four-point fundal accommodation visual score (1 = abnormal, 2 = probably abnormal, 3 = probably normal, and 4 = normal).
RESULTS: Fifty patients had normal SGE, and 22 had delayed SGE. The mean FAR was 0.753 ± 0.151. The fundal accommodation visual score and the FAR were significantly moderately correlated ( ρ = 0.455, P < 0.0001). Receiver operating characteristic curve analysis revealed a FAR cut-off value of 0.669 distinguishing normal from abnormal FAR (area under the curve = 0.873; P < 0.001) with 70% sensitivity and 92.3% specificity. A comparison of the fundal accommodation visual score and FAR revealed highly significant agreement ( P < 0.0001).
CONCLUSION: A fundal radioactivity to total stomach ratio of 0.669 at zero-time image was an appropriate threshold for detecting abnormal fundal accommodation. Given the high correlation between visual scoring and quantitative ratios, visual assessment may be regarded as a quick and dependable way to evaluate fundal accommodation.
METHODS: SGE scintigraphy studies of 72 patients were retrospectively evaluated. Regions of interest (ROIs) were generated on the stomach at different times (0, 30 min, 1, 2, 3, and 4 h), and the corresponding retention percentages were calculated. Counts as geometric means from anterior and posterior ROIs over proximal stomach half and whole stomach at zero-time were used to calculate gastric fundal accommodation ratio (FAR). The radioactivity at the fundus was visually assessed at zero-time image to generate a four-point fundal accommodation visual score (1 = abnormal, 2 = probably abnormal, 3 = probably normal, and 4 = normal).
RESULTS: Fifty patients had normal SGE, and 22 had delayed SGE. The mean FAR was 0.753 ± 0.151. The fundal accommodation visual score and the FAR were significantly moderately correlated ( ρ = 0.455, P < 0.0001). Receiver operating characteristic curve analysis revealed a FAR cut-off value of 0.669 distinguishing normal from abnormal FAR (area under the curve = 0.873; P < 0.001) with 70% sensitivity and 92.3% specificity. A comparison of the fundal accommodation visual score and FAR revealed highly significant agreement ( P < 0.0001).
CONCLUSION: A fundal radioactivity to total stomach ratio of 0.669 at zero-time image was an appropriate threshold for detecting abnormal fundal accommodation. Given the high correlation between visual scoring and quantitative ratios, visual assessment may be regarded as a quick and dependable way to evaluate fundal accommodation.