External Validation of Fully-Automated Infrarenal Maximum Aortic Aneurysm Diameter Measurements in Computed Tomography Angiography Scans Using Artificial Intelligence (PRAEVAorta 2).

Hatzl Johannes, Uhl Christian, Barb Alexandru, Henning Daniel, Fiering Jonathan, El-Sanosy Edris, Cuypers Philippe W M, Böckler Dittmar, Zephyr Study Group Collaborators

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This study investigates the accuracy of fully-automated maximum aortic diameter measurements in abdominal aortic aneurysm (AAA) patients using artificial intelligence software (PRAEVAorta 2, Nurea, Bordeaux, France).

MATERIALS AND

METHODS

This is a multicenter, retrospective validation study using prospectively collected data from the Zenith alpha for aneurysm Repair Registry (ZEPHYR). Automated measurements of PRAEVAorta 2 are compared with measurements of an internationally recognized core laboratory (Syntactx, New York, New York State). The reviewers at the core laboratory were measurement technologists trained to and utilizing established measurement standards, overseen by vascular surgeons and radiologists. The data set comprised 871 computed tomography angiography scans from the ZEPHYR registry with 347 patients who underwent endovascular aneurysm repair (EVAR) with the Zenith Alpha Endovascular Abdominal Graft (Cook Medical, Bloomington, Indiana) in Germany, Belgium, and The Netherlands between 2016 and 2019.

RESULTS

The analysis demonstrated excellent correlation of the measurements (r=0.97) with an intraclass correlation (ICC) of 0.972 (95% confidence interval [CI]=0.968-0.976) across all scans. For preoperative computed tomography (CT) scans, ICC was 0.953 (95% CI=0.941-0.963), and for postoperative scans, ICC was 0.979 (95% CI=0.975-0,983), respectively. In total, 95.4% of measurements were within the clinically acceptable range of 5 mm in absolute difference. In total, 10% of scans demonstrated obvious segmentation errors, mainly due to failure in detecting vessel segments (renal arteries, aortic bifurcation) or due to mis-detecting the outer border of the AAA (duodenum, inferior vena cava, aortic branches) and were excluded from the analysis.

CONCLUSION

In this study, the maximum AAA diameter could be accurately measured fully-automatically by PRAEVAorta 2 (Nurea) in most cases demonstrating that artificial intelligence (AI) software could serve as an important adjunct for research and clinical practice. However, critical review of the generated reports by an experienced observer and cautious use is warranted to identify flawed segmentations.

CLINICAL IMPACT: This multicenter, retrospective validation study assessed the accuracy of fully-automated maximum infrarenal aortic aneurysm diameter measurements. It was demonstrated, that 95.4% of measurements were within the clinically acceptable range of 5 mm in absolute difference, positioning the software as a potential adjunct for clinical practice and research. It is also highlighted however, that critical review of the measurements is obligatory, due to a 10% segmentation error rate.

Для цитирования: Hatzl Johannes, Uhl Christian, Barb Alexandru, Henning Daniel, Fiering Jonathan, El-Sanosy Edris, Cuypers Philippe W M, Böckler Dittmar, Zephyr Study Group Collaborators. External Validation of Fully-Automated Infrarenal Maximum Aortic Aneurysm Diameter Measurements in Computed Tomography Angiography Scans Using Artificial Intelligence (PRAEVAorta 2).. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2024. DOI: 10.1177/15266028241295563.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.