DOI 10.1007/s00392-023-02194-4
Multicenter comparison of transcatheter aortic valve implantation with the self-expanding ACURATE neo2 versus Evolut PRO transcatheter heart valves.
Аннотация
BACKGROUND
New-generation self-expanding transcatheter aortic heart valves (THV) were designed to overcome technical constraints of their preceding generations. We sought to compare the efficacy and safety of the self-expanding ACURATE neo2 (Neo2) versus Evolut PRO (PRO) devices.
METHODS
Seven hundred nine patients undergoing transfemoral transcatheter aortic valve implantation (TAVI) with either Neo2 (n = 496) or PRO (n = 213) were included. Propensity score matching (PSM) was performed to account for differences in baseline characteristics. In-hospital and 30-day clinical outcomes were evaluated according to Valve Academic Research Consortium-3 criteria.
RESULTS
Baseline characteristics were comparable between both groups after PSM (Neo2: n = 155, Evolut Pro: n = 155). Technical success rates were high in both groups (Neo2: 94.8% vs PRO: 97.4%; p = 0.239). Need for permanent pacemaker implantation was less frequent with Neo2 compared with PRO (7.5% vs 20.6%; p = 0.002), whereas major vascular complications were more frequent with Neo2 (Neo2: 11.6% vs PRO: 4.5%; p = 0.022). Intended valve performance at discharge was high in both groups without relevant differences among groups (Neo2: 97.4% vs. 95.3%; p = 0.328).
CONCLUSIONS
Short-term outcomes after TAVI using latest-generation self-expanding THV were excellent, with overall low rates of adverse events. However, Neo2 was associated with lower pacemaker rates and reduced the prevalence of moderate-severe paravalvular leakage. Transprosthetic gradients after TAVI were higher with Neo2 compared with PRO.