ACURATE neo2 versus SAPIEN 3 Ultra for transcatheter aortic valve implantation.

Pellegrini Costanza, Rheude Tobias, Renker Matthias, Wolf Alexander, Wambach Jan Martin, Alvarez-Covarrubias Hector A, Dörr Oliver, Singh Parminder, Charitos Efstratios, Xhepa Erion, Joner Michael, Kim Won-Keun

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

No comparative data exist with the latest generation self-expanding ACURATE neo2 (Neo2) and the balloon-expandable SAPIEN 3 Ultra (Ultra) transcatheter heart valves (THV).

AIMS: We aimed to compare the outcomes after transcatheter aortic valve implantation (TAVI) using the Neo2 and the Ultra THV.

METHODS

A total of 1,356 patients at 4 centres were treated either with the Neo2 (n=608) or the Ultra (n=748). The primary endpoint was device success according to the latest Valve Academic Research Consortium definitions. The association of the THV used and the primary endpoint was assessed using inverse probability treatment weighting (IPTW) and 1:1 propensity score matching (PSM), which identified 472 matched pairs.

Results

After PSM, there were no relevant differences between the groups. While rates of moderate to severe paravalvular leakage (PVL) were overall low (0.6% vs 1.1%; p=0.725), elevated transvalvular gradients (≥20 mmHg) were less frequent with the Neo2 (2.4% vs 7.7%; p<0.001), which translated into a significantly higher rate of device success with the Neo2 compared with the Ultra (91.9% vs 85.0%; p<0.001). Consistently, the Neo2 was associated with higher rates of device success in the IPTW analysis (odds ratio [OR] 1.961, 95% confidence interval [CI]: 1.269-3.031; p=0.002). Rates of mild PVL were significantly lower with the Ultra compared with the Neo2 (20.0% vs 32.8%; p<0.001). Clinical events at 30 days were comparable between the 2 groups.

CONCLUSIONS

Short-term outcomes after TAVI using the Neo2 or Ultra THV were excellent and, overall, comparable. However, transvalvular gradients were lower with the Neo2, which translated into higher rates of device success. Rates of mild PVL were significantly lower with the Ultra THV.

Для цитирования: Pellegrini Costanza, Rheude Tobias, Renker Matthias, Wolf Alexander, Wambach Jan Martin, Alvarez-Covarrubias Hector A, Dörr Oliver, Singh Parminder, Charitos Efstratios, Xhepa Erion, Joner Michael, Kim Won-Keun. ACURATE neo2 versus SAPIEN 3 Ultra for transcatheter aortic valve implantation.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2023. DOI: 10.4244/eij-d-22-00164.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.