DOI 10.1002/jum.14248

Медицинская библиотека · Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine · 2017

Agreement Between an Automated Volume Breast Scanner and Handheld Ultrasound for Diagnostic Breast Examinations.

Barr Richard G, DeVita Robert, Destounis Stamatia, Manzoni Federica, De Silvestri Annalisa, Tinelli Carmine

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To compare the agreement and interobserver variability of diagnostic handheld ultrasound (US) and a single volume on an automated breast volume scanner (ABVS) and to determine whether there was a significant difference if the ABVS was used by a sonographer or mammographic technologist.

METHODS

Ninety patients scheduled for diagnostic US examinations were randomized to either handheld US or the ABVS first. The AVBS was randomized between a sonographer and a mammographic technologist performing the study. The studies were blinded, randomized, and read by 2 radiologists. The lesion with the highest Breast Imaging Reporting and Data System (BI-RADS) score was used in the analysis. Final diagnoses were made by core biopsy or follow-up for 2 years. Lesions included 9 malignant and 81 benign.

RESULTS

The 90 patients had a mean age ± SD of 53.1 ± 16.3 years. The κ value for agreement between the ABVS and handheld US was 0.831 (95% confidence interval, 0.744-0.925), whereas the global agreement for a 7-point BI-RADS score was 0.488 (0.372-0.560). The agreement between the ABVS and handheld US was nearly the same when the ABVS was used by a mammographic technologist (κ = 0.858 [0.723-0.963]) or sonographer (κ = 0.803 [0.596-1.000]; P = .47). The areas under the receiver operating characteristic curves for characterization by the ABVS were 0.91 (0.84-0.96) for reader 1 and 0.91 (0.83-0.96) for reader 2; those for handheld US were 0.91 (0.84-0.96) for reader 1 and 0.83 (0.74-0.90) for reader 2, with no statistical difference. The agreement based on pathologic images was κ = 0.831 (0.718-0.944); for handheld US, κ = 0.795 (0.623-0.967); and for the AVBS, κ = 0.869 (0.725-1.000).

CONCLUSIONS

Performing a single-view diagnostic ABVS examination has good agreement with a handheld diagnostic US workup. There is no difference if the ABVS is used by a sonographer or mammographic technologist.

Для цитирования: Barr Richard G, DeVita Robert, Destounis Stamatia, Manzoni Federica, De Silvestri Annalisa, Tinelli Carmine. Agreement Between an Automated Volume Breast Scanner and Handheld Ultrasound for Diagnostic Breast Examinations.. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2017. DOI: 10.1002/jum.14248.