Jain Ashali, Khalid Maria, Qureshi Muhammad M, Georgian-Smith Dianne, Kaplan Jonah A, Buch Karen, Grinstaff Mark W, Hirsch Ariel E, Hines Neely L, Anderson Stephan W, Gallagher Katherine M, Bates David D B, Bloch B Nicolas
Аннотация
OBJECTIVES: To evaluate breast biopsy marker migration in stereotactic core needle biopsy procedures and identify contributing factors.
METHODS: This retrospective study analyzed 268 stereotactic biopsy markers placed in 263 consecutive patients undergoing stereotactic biopsies using 9G vacuum-assisted devices from August 2010-July 2013. Mammograms were reviewed and factors contributing to marker migration were evaluated. Basic descriptive statistics were calculated and comparisons were performed based on radiographically-confirmed marker migration.
RESULTS: Of the 268 placed stereotactic biopsy markers, 35 (13.1%) migrated ≥1 cm from their biopsy cavity. Range: 1-6 cm; mean (± SD): 2.35 ± 1.22 cm. Of the 35 migrated biopsy markers, 9 (25.7%) migrated ≥3.5 cm. Patient age, biopsy pathology, number of cores, and left versus right breast were not associated with migration status (P> 0.10). Global fatty breast density (P= 0.025) and biopsy in the inner region of breast (P = 0.031) were associated with marker migration. Superior biopsy approach (P= 0.025), locally heterogeneous breast density, and t-shaped biopsy markers (P= 0.035) were significant for no marker migration.
CONCLUSIONS: Multiple factors were found to influence marker migration. An overall migration rate of 13% supports endeavors of research groups actively developing new biopsy marker designs for improved resistance to migration.
KEY POINTS: • Breast biopsy marker migration is documented in 13% of 268 procedures. • Marker migration is affected by physical, biological, and pathological factors. • Breast density, marker shape, needle approach etc. affect migration. • Study demonstrates marker migration prevalence; marker design improvements are needed.
METHODS: This retrospective study analyzed 268 stereotactic biopsy markers placed in 263 consecutive patients undergoing stereotactic biopsies using 9G vacuum-assisted devices from August 2010-July 2013. Mammograms were reviewed and factors contributing to marker migration were evaluated. Basic descriptive statistics were calculated and comparisons were performed based on radiographically-confirmed marker migration.
RESULTS: Of the 268 placed stereotactic biopsy markers, 35 (13.1%) migrated ≥1 cm from their biopsy cavity. Range: 1-6 cm; mean (± SD): 2.35 ± 1.22 cm. Of the 35 migrated biopsy markers, 9 (25.7%) migrated ≥3.5 cm. Patient age, biopsy pathology, number of cores, and left versus right breast were not associated with migration status (P> 0.10). Global fatty breast density (P= 0.025) and biopsy in the inner region of breast (P = 0.031) were associated with marker migration. Superior biopsy approach (P= 0.025), locally heterogeneous breast density, and t-shaped biopsy markers (P= 0.035) were significant for no marker migration.
CONCLUSIONS: Multiple factors were found to influence marker migration. An overall migration rate of 13% supports endeavors of research groups actively developing new biopsy marker designs for improved resistance to migration.
KEY POINTS: • Breast biopsy marker migration is documented in 13% of 268 procedures. • Marker migration is affected by physical, biological, and pathological factors. • Breast density, marker shape, needle approach etc. affect migration. • Study demonstrates marker migration prevalence; marker design improvements are needed.