m+Medelevator.RU
← Поиск публикаций
Heart rhythm · 2017

Relative timing of near-field and far-field electrograms can determine the tachyarrhythmia site of origin.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Caldwell Jane, Gula Lorne, Ali Fariha Sadiq, Miranda Rodrigo I, Abdollah Hoshiar, Baranchuk Adrian, Michael Kevin, Simpson Christopher, Redfearn Damian P

Аннотация

BACKGROUND: Despite improving algorithms, inappropriate shocks for supraventricular tachycardia (SVT) still occur in a significant number of patients with implantable cardioverter-defibrillators (ICDs). This makes the discovery of novel discriminators that use existing ICD hardware an attractive proposition.

OBJECTIVE: We hypothesized that the delay of activation onset from the device-detected, far-field electrogram (EGM) to the near-field, bipole EGM would allow the differentiation of ventricular tachycardias (VTs) from SVTs.

METHODS: Proof of principle was demonstrated by rapid pacing in the right atrium, right ventricle, and left ventricle in healthy patients undergoing atrial fibrillation ablation procedures (n = 17). Using real-life ICD recordings, the equivalent measurements were made in a derivation cohort (n = 26) and cutoff predictive values obtained. Finally, the selected values were validated in a separate group of recordings (n = 82).

RESULTS: In healthy patients, significant differences in the far-field to near-field EGM activation onsets were observed between right atrial (14.7 ± 2.7 ms), right ventricular (36.3 ± 8 ms), and left ventricular (57.8 ± 10.3 ms; P < .001) pacing. In the derivation ICD cohort, the median far-field to near-field onset delay was significantly shorter in SVT (24.5 ms; interquartile range, 15.3-47.5 ms) than in VT (118.5 ms; interquartile range, 102.5-131.5 ms) (P < .001). Using a cutoff of 100 ms in the validation cohort, SVT was successfully discriminated from VT with a sensitivity and specificity of 88% and a negative predictive value of 94.2%.

CONCLUSION: The delay between far-field and near-field EGMs offers a potential new discrimination tool to reduce inappropriate ICD therapies and aid interpretation of single-lead device tracings.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты