DOI 10.1016/j.jcrs.2014.08.043
Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devices.
Аннотация
PURPOSE: To determine the repeatability of keratometry (K) measurements of a Scheimpflug pachymeter (Pentacam), Placido topographer (Eyesys), scanning-slit corneal topographer (Orbscan), partial coherence interferometry (PCI) device (IOLMaster), and Javal manual keratometer with different grades of keratoconus.
SETTING
Noor Eye Hospital, Tehran, Iran.
DESIGN
Retrospective case series.
METHODS
Keratometry was performed first with Scheimpflug pachymetry followed, in order, by Placido topography, scanning-slit corneal topography, PCI, and manual keratometry. Repeatability was examined in groups with a maximum K of less than 50.0 diopters (D), 50.0 to 55.0 D, and more than 55.0 D.
RESULTS
Seventy-eight eyes of 45 keratoconus patients were assessed. In Group 1, repeatability was highest with Scheimpflug pachymetry and lowest with scanning-slit corneal topography (0.36 to 1.24). In Group 2, the intraclass correlation coefficients (ICCs) for maximum K ranged from 0.823 with the scanning-slit corneal topography to 0.974 with Scheimpflug pachymetry. The repeatability index for minimum K (0.53 to 2.11) and maximum K (0.60 to 1.92) in this group was highest with Scheimpflug pachymetry and with for scanning-slit corneal topography. In Group 3, the ICCs for minimum K and maximum K ranged from 0.890 to 0.990, and the repeatability index for minimum K varied between 1.66 with Scheimpflug pachymetry to 2.98 with Placido topography; for maximum K, the index was from 2.15 with PCI to 2.81 with the manual keratometer.
CONCLUSIONS
In mild keratoconus, the 5 devices had acceptable repeatability in K readings. In cases with a maximum K reading greater than 55.0 D, all devices had reduced repeatability as a result of measurement errors; thus, measurements might not be so reliable.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.