Repeatability of corneal phase retardation measurements by polarization-sensitive optical coherence tomography.

Fukuda Shinichi, Kishino Genichiro, Hoshi Sujin, Beheregaray Simone, Ueno Yuta, Fukuda Masahiro, Kasaragod Deepa, Yasuno Yoshiaki, Oshika Tetsuro

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Polarization-sensitive optical coherence tomography (PS-OCT) can evaluate internal tissue structures of the cornea, such as collagen fibers, by phase retardation measurement. In this study, we assessed the repeatability of corneal phase retardation measurements using anterior segment PS-OCT.

METHODS

A total of 173 eyes of 173 patients were measured using PS-OCT. In total, 58 eyes of young subjects with normal corneas, 28 eyes of old subjects with normal corneas, 26 eyes with corneal dystrophy or degeneration, 37 eyes with corneal transplantation, and 24 eyes with keratoconus were evaluated. The 3-mm diameter average of en face phase retardation of the posterior corneal surface was examined using PS-OCT. To evaluate the repeatability, intraclass correlation coefficients (ICCs) were calculated for intraobserver repeatability and interobserver repeatability analysis.

RESULTS

Polarization-sensitive OCT showed good repeatability for corneal measurements. Intraclass correlation coefficients of intraobserver and interobserver repeatability of all the subjects were 0.989 and 0.980, respectively. Intraclass correlation coefficients of the intraobserver for each group, that is, young and old subjects with normal cornea, cornea dystrophy/degeneration, corneal transplantation, and keratoconus, were 0.961, 0.975, 0.984, 0.978, and 0.996, respectively. Interobserver ICCs for the above-mentioned respective groups were 0.952, 0.964, 0.988, 0.959, and 0.975, respectively.

CONCLUSIONS

Polarization-sensitive OCT showed good repeatability for phase retardation measurements of central corneas not only for normal corneas, but also for various diseased corneas. Polarization-sensitive OCT might be useful for evaluating corneal phase retardation, which is one of the parameters that defines birefringence.

Для цитирования: Fukuda Shinichi, Kishino Genichiro, Hoshi Sujin, Beheregaray Simone, Ueno Yuta, Fukuda Masahiro, Kasaragod Deepa, Yasuno Yoshiaki, Oshika Tetsuro. Repeatability of corneal phase retardation measurements by polarization-sensitive optical coherence tomography.. Investigative ophthalmology & visual science. 2015. DOI: 10.1167/iovs.14-16327.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.