DOI 10.1016/j.jcrs.2014.06.043

Медицинская библиотека · Journal of cataract and refractive surgery · 2015

Corneal epithelial thickness mapping using Fourier-domain optical coherence tomography for detection of form fruste keratoconus.

Temstet Cyril, Sandali Otman, Bouheraoua Nacim, Hamiche Taous, Galan Alice, El Sanharawi Mohamed, Basli Elena, Laroche Laurent, Borderie Vincent

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine whether optical coherence tomography (OCT) epithelial mapping can improve the detection of form fruste keratoconus.

SETTING

French National Eye Hospital, Paris 6 Pierre & Marie Curie University, Paris, France.

DESIGN

Retrospective comparative study.

METHODS

Eyes with normal corneas, form fruste keratoconus, moderate keratoconus, or severe keratoconus were assessed using Fourier-domain OCT (RTVue 5.5), scanning-slit corneal topography (Orbscan IIz), and rotating Scheimpflug camera (Pentacam Comprehensive Eye Scanner). Several parameters provided by the software or derived from elevation maps, OCT pachymetric maps, and OCT epithelium parameters were evaluated and compared between the 4 groups.

RESULTS

The study involved 145 eyes. There were no significant differences in the keratometry (K) value, inferior-superior value, keratoconus index, central K index, and topographic keratoconus classification indices between the form fruste keratoconus group and the control group (P > .05). Form fruste keratoconic corneas had less epithelial thickness in the thinnest corneal zone than normal corneas, and greater epithelial thickness in the thinnest corneal zone than keratoconic corneas (P < .005). The epithelial thickness in the thinnest corneal zone in form fruste corneas was located inferiorly (P < .005) and corresponded with the zone of minimum epithelial thickness and maximum posterior elevation (P < .005). The receiver operating characteristic curve analysis showed good overall predictive accuracy of the epithelial thickness in the thinnest corneal zone, with a 52 μm threshold value for discriminating form fruste keratoconic corneas from normal corneas.

CONCLUSIONS

The epithelial thickness in the thinnest corneal zone and its location provided by the OCT epithelial mapping might be useful for the early diagnosis of form fruste keratoconus.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Temstet Cyril, Sandali Otman, Bouheraoua Nacim, Hamiche Taous, Galan Alice, El Sanharawi Mohamed, Basli Elena, Laroche Laurent, Borderie Vincent. Corneal epithelial thickness mapping using Fourier-domain optical coherence tomography for detection of form fruste keratoconus.. Journal of cataract and refractive surgery. 2015. DOI: 10.1016/j.jcrs.2014.06.043.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.