[The study of corneal epithelial change in keratoconic eyes with Fourier-domain optical coherence tomographic pachymetry].

Zhao Yujin, Hong Jiaxu, Wang Fei, Cui Xinhan, Yang Yujing, Zhu Xiaoyu, Wu Dan, Wei Anji, Chen Yan, Wu Suqian, Huang Feifei, Xu Jianjiang

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the corneal epithelial thickness profile in keratoconic and normal eyes with advanced Fourier-domain optical coherence tomography (OCT) and to evaluate the diagnostic efficiency of pachymetry parameters.

METHODS

Cross-sectional observational study. Twenty-five consecutive keratoconus patients (34 eyes) and 41 normal volunteers (41 eyes) in Eye & ENT Hospital of Fudan University between June 1 and July 31, 2013 were enrolled.Observations of corneal epithelial thickness and corneal thickness profiles were taken from all subjects. The epithelial pachymetry parameters were superior, inferior, inferior-superior (I-S), minimum (Min), maximum (Max), Min-Max, and central. The corneal pachymetry parameters were inferotemporal-superonasal (IT-SN), I-S, Min, Min-median, Min-Max, and central.Statistical analysis methods included Chi-square test, independent two-sample t-test, and Wilcoxon rank sum test. The diagnostic efficiency of all parameters were analyzed by receiver operating characteristic (ROC) curves and Youden index.

RESULTS

The inferior, I-S, Min, Min-Max, and central epithelial thickness averages were (50.5 ± 4.8) µm, (-2.7 ± 7.5) µm, (35.9 ± 10.7) µm, (-33.6 ± 18.9) µm, and (50.1 ± 6.8) µm in keratoconic eyes.In normal eyes, they were (54.8 ± 2.5) µm, (0.9 ± 1.3) µm, (51.4 ± 2.6) µm, (-5.7 ± 1.8) µm, and (54.7 ± 2.2) µm. Among the epithelial pachymetry parameters, Min-Max and Min showed high diagnostic efficiency for keratoconus (AUC = 0.967, 0.968; Youden index = 0.882, 0.829); inferior, I-S, and central parameters showed moderate efficiency (AUC = 0.794, 0.728, 0.715; Youden index = 0.461, 0.564, 0.565). The IT-SN, I-S, Min, Min-median, Min-Max, and central corneal thickness averages were (-82.8 ± 65.2) µm, (-72.5 ± 55.6) µm, (401.2 ± 83.8) µm, (-83.4 ± 48.2) µm, (-169.7 ± 99.4) µm, and (453.4 ± 58.6) µm in keratoconic eyes.In normal eyes, these parameters were respectively (-23.7 ± 18.5) µm, (-20.5 ± 19.8) µm, (525.4 ± 28.1) µm, (-23.1 ± 4.8) µm, (-59.2 ± 23.3) µm, and (533.4 ± 28.8) µm. Min, Min-median, and Min-Max had preferable diagnostic efficiency (AUC = 0.952, 0.938, 0.924; Youden index = 0.784, 0.829, 0.721); central, IT-SN, and I-S parameters showed moderate efficiency (AUC = 0.891, 0.877, 0.862; Youden index = 0.643, 0.707, 0.580).

CONCLUSIONS

In observation with Fourier-domain OCT, keratoconus was characterized by central and inferior epithelial thinning, increased epithelial thickness variation, central cornea thinning, increase in asymmetry, and gaps between corneal IT-SN and I-S thickness.Fourier-domain OCT was confirmed to have high speed, high resolution, and noninvasion. The pachymetry parameters showed specific strengths in diagnosis of keratoconus, especially epithelial pachymetry parameters of Min and Min-Max, and corneal pachymetry parameters of Min, Min-median and Min-Max.

Для цитирования: Zhao Yujin, Hong Jiaxu, Wang Fei, Cui Xinhan, Yang Yujing, Zhu Xiaoyu, Wu Dan, Wei Anji, Chen Yan, Wu Suqian, Huang Feifei, Xu Jianjiang. [The study of corneal epithelial change in keratoconic eyes with Fourier-domain optical coherence tomographic pachymetry].. [Zhonghua yan ke za zhi] Chinese journal of ophthalmology. 2014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.