Comparative analysis of corneal measurements obtained from a Scheimpflug camera and an integrated Placido-optical coherence tomography device in normal and keratoconic eyes.

Viswanathan Deepa, Kumar Nikhil L, Males John J, Graham Stuart L

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the agreement between a Scheimpflug camera (Pentacam) and a combined Placido-optical coherence tomography device (Visante OMNI) in measuring corneal curvature, thickness and elevation values in normal and keratoconic eyes.

METHODS

Corneal measurements of 110 normal eyes (one eye per subject) and 70 keratoconic eyes were obtained from both devices and compared. Agreement was determined using the Bland-Altman analysis 95% limits of agreement (LoA).

RESULTS

The Pentacam measured significantly greater keratometry readings in the flattest (K1) and steepest meridians (K2) in normal and keratoconic eyes. The 95% LoA in normal eyes were -0.32 to 0.59 dioptres (D) (K1) and -0.41 to 0.74 D (K2). In keratoconic eyes, the 95% LoA were -1.35 to 1.92 D (K1) and -1.38 to 1.99 D (K2). The Pentacam recorded significantly higher central corneal thickness (CCT) values in both groups of eyes. The 95% LoA were -4.31 to 39.89 microns (μ) and -12.92 to 41.35 μ in normal and keratoconic eyes, respectively. Pentacam anterior and posterior corneal elevations were significantly greater in both groups of eyes. The devices demonstrated excellent repeatability and reproducibility for corneal curvature and thickness but not elevation measurements.

CONCLUSIONS

The Pentacam measured significantly greater corneal curvature, thickness and elevation values compared to the Visante OMNI in normal and keratoconic eyes. The devices agree moderately for anterior corneal elevations in normal eyes and do not appear to be interchangeable for corneal measurements in clinical practice.

Для цитирования: Viswanathan Deepa, Kumar Nikhil L, Males John J, Graham Stuart L. Comparative analysis of corneal measurements obtained from a Scheimpflug camera and an integrated Placido-optical coherence tomography device in normal and keratoconic eyes.. Acta ophthalmologica. 2014. DOI: 10.1111/aos.12622.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.