A precise, sensitive and stable LC-MSMS method for detection of picomolar levels of serum aldosterone.

Lie Margrete, Thorstensen Ketil

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Accurate quantification of aldosterone is essential in the diagnosis of primary aldosteronism. Liquid chromatography-tandem mass spectrometry (LC-MSMS) analysis is increasingly being used to improve analytical sensitivity and specificity, since this technology reduces most of the interferences observed with immunological methods.

METHODS

Serum samples with d7-aldosterone as internal standard were extracted with methyl tert-butyl ether, using liquid-liquid extraction (LLE). Chromatographic separation was performed on a C18 reverse phase column with a methanol-water gradient containing ammonium fluoride. Aldosterone detection was performed on an Agilent 6490 triple quadrupole using electro spray ionisation in positive mode.

RESULTS

Multiple reaction monitoring transitions were m/z 361.2-315.1 for aldosterone, and 368.5-323.3 for d7-aldosterone. Chromatographic retention time was 2.7 min. The method's total CVs at aldosterone concentrations of 45.4 and 1080 pmol/L were 7.0% and 4.8%, respectively. The intra-assay CVs at concentrations of 60.0 and 637 pmol/L were 4.0% and 2.6%, respectively. The method's LOQ and LOD were 10 and 5 pmol/L, respectively, demonstrating an excellent analytical sensitivity. The upper limit of quantification was set to 5000 pmol/L, corresponding to the highest calibrator concentration. The long-term stability of the method was evident from repeated measurements of external control pools from UKNEQAS over a period of about 3 years, showing CVs between 2.0 and 7.0%.

CONCLUSIONS

We have described a precise, sensitive and stable LC-MSMS method for the measurement of serum aldosterone. In addition, due to the use of LLE and a short LC-column, the method is simple to perform, with a short chromatographic run time.

Для цитирования: Lie Margrete, Thorstensen Ketil. A precise, sensitive and stable LC-MSMS method for detection of picomolar levels of serum aldosterone.. Scandinavian journal of clinical and laboratory investigation. 2018. DOI: 10.1080/00365513.2018.1480060.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.