Concurrent validity of a novel digital dynamometerfor assessing performance on the Cranio-Cervical Flexion Test.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Barbero Marco, Lonati Daniele, Griseri Marco, Cescon Corrado, Bonafine Alessandro, Negrisolo Tiziano, Falzone Emanuele, Pedrucci Samuel, Cattrysse Erik, Falla Deborah

Аннотация

INTRODUCTION: The Cranio-Cervical Flexion Test (CCFT) is a validated clinical test used to evaluate deep cervical flexor muscle function. A pressure biofeedback unit (PBU) has traditionally been the reference device. Recently, a new portable digital device known as the Neuromuscular Cranio-cervical Device (NOD) has been developed as a potential alternative.

METHODS: In this cross-sectional concurrent validation study, 88 participants were recruited from four outpatient physiotherapy practices, including individuals with current neck pain and individuals with a previous history of neck pain. All participants performed the CCFT using both devices. The PBU was inflated to a baseline pressure of 20 mmHg, with the participant then performing C-CF to increase the pressure in increments of 2 mmHg over five progressive stages, while the NOD was calibrated to corresponding force targets. Agreement and concurrent validity between the devices were assessed, and C-CF range of motion (ROM) was analyzed using linear mixed-effects models.

RESULTS: Agreement between CCFT performance scores obtained with the NOD and the PBU was 94.3%, with a very strong positive correlation (Spearman's ρ = 0.930, p < 0.001). The NOD measured slightly greater C-CF ROM than the PBU (mean difference 0.68°), with significant differences at lower CCFT stages that converged at higher stages of the test.

CONCLUSION: These findings demonstrate strong concurrent validity between the NOD and the PBU, supporting the NOD as a valid digital alternative for assessment of the CCFT.