DOI 10.1016/j.ejrad.2023.110736

Медицинская библиотека · European journal of radiology · 2023

Reducing both radiation and contrast doses for overweight patients in coronary CT angiography with 80-kVp and deep learning image reconstruction.

Li Wanjiang, Lu Haiyan, Wen Yuting, Zhou Minggang, Shuai Tao, You Yongchun, Zhao Jin, Liao Kai, Lu Chunyan, Li Jianying, Li Zhenlin, Diao Kaiyue, He Yong

Рандомизированное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the use of an 80-kVp tube voltage combined with a deep learning image reconstruction (DLIR) algorithm in coronary CT angiography (CCTA) for overweight patients to reduce radiation and contrast doses in comparison with the 120-kVp protocol and adaptive statistical iterative reconstruction (ASIR-V).

METHODS

One hundred consecutive CCTA patients were prospectively enrolled and randomly divided into a low-dose group (n = 50) with 80-kVp, smart mA for noise index (NI) of 36 HU, contrast dose rate of 18 mgI/kg/s and DLIR and 60 % ASIR-V and a standard-dose group (n = 50) with 120-kVp, smart mA for NI of 25 HU, contrast dose rate of 32 mgI/kg/s and 60 % ASIR-V. The radiation and contrast dose, subjective image quality score, attenuation values, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were compared.

RESULTS

The low-dose group achieved a significant reduction in the effective radiation dose (1.01 ± 0.45 mSv vs 1.85 ± 0.40 mSv, P < 0.001) and contrast dose (33.69 ± 3.87 mL vs 59.11 ± 5.60 mL, P < 0.001) compared to the standard-dose group. The low-dose group with DLIR presented similar enhancement but lower noise, higher SNR and CNR and higher subjective quality scores than the standard-dose group. Moreover, the same patient comparison in the low-dose group between different reconstructions showed that DLIR images had slightly and consistently higher CT values in small vessels, indicating better defined vessels, much lower image noise, higher SNR and CNR and higher subjective quality scores than ASIR-V images (all P < 0.001).

CONCLUSIONS

The application of 80-kVp and DLIR allows for significant radiation and dose reduction while further improving image quality in CCTA for overweight patients.

Для цитирования: Li Wanjiang, Lu Haiyan, Wen Yuting, Zhou Minggang, Shuai Tao, You Yongchun, Zhao Jin, Liao Kai, Lu Chunyan, Li Jianying, Li Zhenlin, Diao Kaiyue, He Yong. Reducing both radiation and contrast doses for overweight patients in coronary CT angiography with 80-kVp and deep learning image reconstruction.. European journal of radiology. 2023. DOI: 10.1016/j.ejrad.2023.110736.